Event Text Entries

[116915233] Investigation including root cause analysis is in progress. A supplemental mdr will be filed as necessary in accordance with 21 cfr 803. 56 when additional reportable information becomes available (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[116915234] A customer reported that the microscope optics and head had issues. Additional information was requested but not received.
Patient Sequence No: 1, Text Type: D, B5


[129970985] The system was examined and the reported event was confirmed. The yoke set screw was replaced to resolve the issue. The system was tested and found to meet product specifications. A review of the customer? S complaint history for the last 24 months did not show any previous complaints of this kind against the system. The system was manufactured on february 20, 2017. Based on assessment, the product met specifications at the time of release. The root cause of the reported event can be attributed to a physically damaged yoke set screw. However, how or when the yoke set screw became nonconforming is unable to be determined conclusively. The manufacturer internal reference number is: (b)(4).
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2028159-2018-01683
MDR Report Key7771582
Date Received2018-08-10
Date of Report2018-12-11
Date Mfgr Received2018-11-26
Device Manufacturer Date2017-02-20
Date Added to Maude2018-08-10
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. BRYAN BLAKE
Manufacturer Street6201 SOUTH FREEWAY MAIL STOP AB2-6
Manufacturer CityFORT WORTH TX 76134
Manufacturer CountryUS
Manufacturer Postal76134
Manufacturer Phone8176152230
Manufacturer G1ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER
Manufacturer Street15800 ALTON PARKWAY
Manufacturer CityIRVINE CA 92618
Manufacturer CountryUS
Manufacturer Postal Code92618
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report0

Device Details

Brand NameLUXOR OPHTHALMIC MICROSCOPES
Generic NameMICROSCOPE, OPERATING & ACC., AC-POWERED, OPHTHALMIC
Product CodeHRM
Date Received2018-08-10
Model NumberLX3
Catalog Number8065752560
Lot NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER
Manufacturer Address15800 ALTON PARKWAY IRVINE CA 92618 US 92618


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2018-08-10

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.