MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2018-08-10 for EMAG? SYSTEM 4704256 manufactured by Biomerieux Sa.
[116718349]
A customer in (b)(6) notified biom? Rieux of potential contamination associated with emag? System (reference 470456). The customer reported that there was a potential contamination on a sample adjacent to an uls error sample on the emag system. Sample results on (b)(6) 2018 for batch z109gpbnm at 9:32 am left section: - vessel position 1. 2 skin swab id = (b)(6) / error 1076 ultrasonic reading below threshold => not extracted. - vessel position 1. 3 csf id = (b)(6) pcr amplification result altona: (b)(6). A second run extraction on the same two samples performed on (b)(6) 2018 at 15:38 vessel 1 position 1 and 2 no uls error on the skin sample altona amplification: (b)(6). Id = (b)(6) csf negative. There was no impact to the patient due to this issue. However, there was a delay in reporting results greater than 24 hours. An internal biom? Rieux investigation will be initiated.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 3002769706-2018-00140 |
MDR Report Key | 7771781 |
Report Source | FOREIGN,HEALTH PROFESSIONAL,U |
Date Received | 2018-08-10 |
Date of Report | 2018-11-05 |
Date Mfgr Received | 2018-10-10 |
Date Added to Maude | 2018-08-10 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Reporter Occupation | OTHER HEALTH CARE PROFESSIONAL |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | MS. TIFFANY HALL |
Manufacturer Street | 100 RODOLPHE STREET |
Manufacturer City | DURHAM NC 27712 |
Manufacturer Country | US |
Manufacturer Postal | 27712 |
Manufacturer G1 | BIOMERIEUX SA |
Manufacturer Street | VIA DI CAMPIGLIANO 58 - LOC. PONTE A EMA |
Manufacturer City | 50012 BAGNO A RIPOLI, |
Manufacturer Country | IT |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | EMAG? SYSTEM |
Generic Name | EMAG? SYSTEM |
Product Code | JJH |
Date Received | 2018-08-10 |
Model Number | 4704256 |
Device Availability | N |
Device Age | DA |
Device Eval'ed by Mfgr | R |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | BIOMERIEUX SA |
Manufacturer Address | VIA DI CAMPIGLIANO 58 - LOC. PONTE A EMA 50012 BAGNO A RIPOLI, IT |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2018-08-10 |