EMAG? SYSTEM 4704256

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2018-08-10 for EMAG? SYSTEM 4704256 manufactured by Biomerieux Sa.

Event Text Entries

[116718349] A customer in (b)(6) notified biom? Rieux of potential contamination associated with emag? System (reference 470456). The customer reported that there was a potential contamination on a sample adjacent to an uls error sample on the emag system. Sample results on (b)(6) 2018 for batch z109gpbnm at 9:32 am left section: - vessel position 1. 2 skin swab id = (b)(6) / error 1076 ultrasonic reading below threshold => not extracted. - vessel position 1. 3 csf id = (b)(6) pcr amplification result altona: (b)(6). A second run extraction on the same two samples performed on (b)(6) 2018 at 15:38 vessel 1 position 1 and 2 no uls error on the skin sample altona amplification: (b)(6). Id = (b)(6) csf negative. There was no impact to the patient due to this issue. However, there was a delay in reporting results greater than 24 hours. An internal biom? Rieux investigation will be initiated.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3002769706-2018-00140
MDR Report Key7771781
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2018-08-10
Date of Report2018-11-05
Date Mfgr Received2018-10-10
Date Added to Maude2018-08-10
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. TIFFANY HALL
Manufacturer Street100 RODOLPHE STREET
Manufacturer CityDURHAM NC 27712
Manufacturer CountryUS
Manufacturer Postal27712
Manufacturer G1BIOMERIEUX SA
Manufacturer StreetVIA DI CAMPIGLIANO 58 - LOC. PONTE A EMA
Manufacturer City50012 BAGNO A RIPOLI,
Manufacturer CountryIT
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameEMAG? SYSTEM
Generic NameEMAG? SYSTEM
Product CodeJJH
Date Received2018-08-10
Model Number4704256
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBIOMERIEUX SA
Manufacturer AddressVIA DI CAMPIGLIANO 58 - LOC. PONTE A EMA 50012 BAGNO A RIPOLI, IT


Patients

Patient NumberTreatmentOutcomeDate
10 2018-08-10

© 2024 FDA.report