SYMPHION? TISSUE REMOVAL SYSTEM REFURBISHED CONTROLLER FG-0612-R

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2018-08-10 for SYMPHION? TISSUE REMOVAL SYSTEM REFURBISHED CONTROLLER FG-0612-R manufactured by Boston Scientific - San Jose.

Event Text Entries

[116867748] The investigation revealed that corrosion was found on the motor board near the hand piece motor circuit. It appeared saline had dripped into the unit where the top cover meets the front bezel and created an electrical short on the main board. The short activated the motor circuits and thus the resecting device triggered by itself. Given the event description and condition of the returned device, the cause was traced back to component failure. A review of the device history record (dhr) confirmed that the device met all material, assembly, and product specifications.
Patient Sequence No: 1, Text Type: N, H10


[116867749] It was reported to boston scientific corporation on (b)(6) 2018 that a symphion controller was used during a procedure performed on (b)(6) 2018. According to the complainant, during the procedure and outside the patient, the symphion resecting device started resecting immediately after connecting it into the symphion controller. The resecting device was inserted through the endoscope while the resecting device was activated in order to complete the procedure. There were no patient complications reported as a result of this event. The patient condition at the conclusion of the procedure was reported to be fine.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3005099803-2018-02536
MDR Report Key7773665
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2018-08-10
Date of Report2018-07-20
Date of Event2018-07-20
Date Mfgr Received2018-07-20
Device Manufacturer Date2014-10-30
Date Added to Maude2018-08-10
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactEMP. NANCY CUTINO
Manufacturer Street100 BOSTON SCIENTIFIC WAY
Manufacturer CityMARLBOROUGH MA 01752
Manufacturer CountryUS
Manufacturer Postal01752
Manufacturer Phone5086834000
Manufacturer G1BOSTON SCIENTIFIC - SAN JOSE
Manufacturer Street150 BAYTECH DRIVE
Manufacturer CitySAN JOSE CA 95134
Manufacturer CountryUS
Manufacturer Postal Code95134
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSYMPHION? TISSUE REMOVAL SYSTEM REFURBISHED CONTROLLER
Generic NameINSUFFLATOR, HYSTEROSCOPIC, FLUID, CLOSED-LOOP RECIRCULATION WITH CUTTER-COAGULA
Product CodePGT
Date Received2018-08-10
Returned To Mfg2018-07-27
Model NumberFG-0612-R
Lot NumberMC-0069
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBOSTON SCIENTIFIC - SAN JOSE
Manufacturer Address150 BAYTECH DRIVE SAN JOSE CA 95134 US 95134


Patients

Patient NumberTreatmentOutcomeDate
10 2018-08-10

© 2026 FDA.report