AIRLIFE? INFANT SKIN TEMPERATURE PROBE, REUSABLE (GIRAFFE?, PANDA?) 2075796-001

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a distributor,health profession report with the FDA on 2018-08-13 for AIRLIFE? INFANT SKIN TEMPERATURE PROBE, REUSABLE (GIRAFFE?, PANDA?) 2075796-001 manufactured by Vyaire Medical.

Event Text Entries

[116960638] At this time, vyaire has not received the suspect device/component for evaluation. If additional information is received a supplemental report will be submitted.
Patient Sequence No: 1, Text Type: N, H10


[116960639] Customer reported that new temp probes are affecting the ability of the incubators at thermal regulation; while the temp probe is attached to the patient, the inflexibility of the cable causes it to lift off of the patient's skin even when covered with the appropriate heat reflecting patch; once the probe has lifted away from the skin the giraffe omnibed receives an inaccurate temp and tries to compensate by increasing the temperature in the bed. This in turn overheats the infant which then takes some time for the infant to recover this incidence. The area of the cable closest to the temp probe is stiff and rigid.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3005934741-2018-00001
MDR Report Key7778834
Report SourceDISTRIBUTOR,HEALTH PROFESSION
Date Received2018-08-13
Date of Report2018-06-26
Date of Event2018-06-26
Date Mfgr Received2018-06-26
Date Added to Maude2018-08-13
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMINDY FABER
Manufacturer Street26125 NORTH RIVERWOODS BLVD
Manufacturer CityMETTAWA IL 60045
Manufacturer CountryUS
Manufacturer Postal60045
Manufacturer Phone8727570116
Manufacturer G1THERMOMETRICS DE MEXICO S.A. DE C.V
Manufacturer StreetCALLE SIETE NORTE NO 110 CIUDAD INDUSTRIAL OTAY
Manufacturer CityTIJUANA, 22444
Manufacturer CountryMX
Manufacturer Postal Code22444
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameAIRLIFE? INFANT SKIN TEMPERATURE PROBE, REUSABLE (GIRAFFE?, PANDA?)
Generic NameCLINICAL ELECTRONIC THERMOMETER
Product CodeFMZ
Date Received2018-08-13
Catalog Number2075796-001
Lot NumberTHM180404
OperatorHEALTH PROFESSIONAL
Device Availability*
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerVYAIRE MEDICAL
Manufacturer Address26125 NORTH RIVERWOODS BLVD METTAWA IL 60045 US 60045