GORE VIATORR? TIPS ENDOPROSTHESIS

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-08-13 for GORE VIATORR? TIPS ENDOPROSTHESIS manufactured by W.l. Gore & Associates.

Event Text Entries

[116996732] The gore? Viatorr? Tips endoprosthesis instructions for use precaution section notes the graft-lined portion of the gore? Viatorr? Tips endoprosthesis should completely cover the intrahepatic tract, preferably to the ostium of the hepatic vein at the inferior vena cava. Additionally, thrombosis is listed as a possible adverse event in the gore? Viatorr? Tips endoprosthesis instructions for use. Regnaulta d, d? Alterochea l, nicolasa c, dujardinb f, ayoubc j, perarnau jm. Ten-year experience of transjugular intrahepatic portosystemic shunt for noncirrhotic portal hypertension. European journal of gastroenterology & hepatology 2018; 30:557? 562.
Patient Sequence No: 1, Text Type: N, H10


[116996733] This information was received through literature article "ten-year experience of transjugular intrahepatic portosystemic shunt for noncirrhotic portal hypertension" published in the european journal of gastroenterology and hepatology, may 2018. The study is a monocentric retrospective study concerning patients treated by tips between 2004 and 2015 for complications of noncirrhotic portal hypertension. The article reports an early complication (< 30 days) with a gore? Viatorr? Tips endoprosthesis. An asymptomatic early thrombosis was detected via ultrasound on day five following implant. It was further reported that too short a stent was initially placed requiring a new and longer 1 cm stent.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3007284313-2018-00225
MDR Report Key7779221
Date Received2018-08-13
Date of Report2018-07-18
Date of Event2018-05-15
Date Added to Maude2018-08-13
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMARCI STEWART
Manufacturer Street1500 N. 4TH STREET
Manufacturer CityFLAGSTAFF AZ
Manufacturer Phone9285263030
Manufacturer G1MEDICAL PHOENIX 1 B/P
Manufacturer Street32360 N. NORTH VALLEY PARKWAY
Manufacturer CityPHOENIX AZ 85085
Manufacturer CountryUS
Manufacturer Postal Code85085
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameGORE VIATORR? TIPS ENDOPROSTHESIS
Generic NameSHUNT, PORTOSYSTEMIC, ENDOPROSTHESIS
Product CodeMIR
Date Received2018-08-13
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerW.L. GORE & ASSOCIATES
Manufacturer AddressFLAGSTAFF AZ


Patients

Patient NumberTreatmentOutcomeDate
101. Other; 2. Required No Informationntervention 2018-08-13

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.