ALPHA OMEGA Standard STR-020121-10

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-08-14 for ALPHA OMEGA Standard STR-020121-10 manufactured by Alpha Omega Engineering Ltd..

Event Text Entries

[117129721] An incident was reported by "department of neurosurgery, (b)(6) university" in (b)(6) (the customer), which uses alpha omega engineering's (aoe) neuro omega system for navigation during neurosurgery for placement a dbs electrode. Aoe's cannula p/n-str-007721-10, with length 177 mm, is normally used for this neurosurgery procedure. In the reported incident, a longer identical cannula was mistakenly used (p/n- str-020121-10, with length 201. 5 mm), resulting a dipper insertion, before the mistake was realized by the customer. The two types of cannulas are used by the customer's for different head frames. The label on the cannula clearly identifies the cannula type (by part number, length and diameter). No injury medical intervention to the patient was reported by customer.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9615126-2018-00002
MDR Report Key7780725
Date Received2018-08-14
Date of Report2018-08-07
Date of Event2018-07-24
Date Mfgr Received2018-07-24
Device Manufacturer Date2017-10-25
Date Added to Maude2018-08-14
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMRS. MAYSANA MOUSA
Manufacturer StreetMOUNT PRECIPICE
Manufacturer CityNAZARETH, 1612102
Manufacturer CountryIS
Manufacturer Postal1612102
Manufacturer G1ALPHA OMEGA ENGINEERING LTD.
Manufacturer StreetMOUNT PRECIPICE
Manufacturer CityNAZARETH, 1612102
Manufacturer CountryIS
Manufacturer Postal Code1612102
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameALPHA OMEGA
Generic NameSTERILE DISPOSABLE CANNULA FOR MER
Product CodeGZL
Date Received2018-08-14
Model NumberStandard
Catalog NumberSTR-020121-10
Lot Number29713
Device Expiration Date2019-10-25
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerALPHA OMEGA ENGINEERING LTD.
Manufacturer AddressMOUNT PRECIPICE NAZARETH, 1612102 IS 1612102


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2018-08-14

© 2026 FDA.report