SINGLE FIBER OPTIC CABLE, 10FT, PINK FC1SP

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2018-08-15 for SINGLE FIBER OPTIC CABLE, 10FT, PINK FC1SP manufactured by Invuity, Inc.

Event Text Entries

[117448187] During evaluation, it was noted that broken fibers were present throughout the cable. Additionally, the cable exhibited a concentrated number of broken fibers which indicates mishandling of the cable. The concentration of broken fibers led to an elevated temperature. Invuity's ifu clearly states to handle the cables with care and not to coil the cables tighter than six (6) inches.
Patient Sequence No: 1, Text Type: N, H10


[117448188] It was reported during a hip replacement procedure, the fiber optic cable may have been placed on the drape of the patient. During use, the temperature of the fiber optic cord felt hot and left a burn mark on the drape. The case was completed without incident or product malfunction. There was no patient injury involved.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3011394215-2018-00006
MDR Report Key7784680
Report SourceHEALTH PROFESSIONAL
Date Received2018-08-15
Date of Report2018-07-17
Date of Event2018-07-17
Date Mfgr Received2018-07-17
Date Added to Maude2018-08-15
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR JOHN KANG
Manufacturer Street444 DE HARO ST
Manufacturer CitySAN FRANCISCO CA 94107
Manufacturer CountryUS
Manufacturer Postal94107
Manufacturer Phone4156552160
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSINGLE FIBER OPTIC CABLE, 10FT, PINK
Generic NameSINGLE FIBER OPTIC CABLE, 10FT, PINK
Product CodeFST
Date Received2018-08-15
Returned To Mfg2018-07-23
Model NumberFC1SP
Catalog NumberFC1SP
Lot NumberUNKNOWN
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerINVUITY, INC
Manufacturer Address444 DE HARO ST SAN FRANCISCO CA 94107 US 94107


Patients

Patient NumberTreatmentOutcomeDate
10 2018-08-15

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.