STREAMLAB

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative report with the FDA on 2018-08-15 for STREAMLAB manufactured by Siemens Healthcare Diagnostics Inc..

Event Text Entries

[117299002] Siemens was notified by the original equipment manufacturer for streamlab that a pacemaker safety label (implanted cardiac device) may be missing from certain streamlab systems. The potentially impacted streamlab modules are the core unit / lynx with input output module (iom) and refrigerated storage module. Siemens was further informed that the minimum safety distance from devices emitting a magnetic field should be increased to 200 millimeters (7. 87 inches). Siemens has issued an urgent medical device correction titled aptio? Automation streamlab? Automation some modules missing pacemaker warning labels ((b)(4)) and an urgent field safety notification titled aptio? Automation streamlab? Automation some modules missing pacemaker warning labels ((b)(4)) to all affected customers.
Patient Sequence No: 1, Text Type: N, H10


[117299003] Siemens was notified by the original equipment manufacturer (oem) for sreamlab that an ongoing investigation has determined that a pacemaker safety label (implanted cardiac device) may be missing from certain streamlab systems. Siemens was further informed that the minimum safety distance from devices emitting a magnetic field should be increased to 200 millimeters (7. 87 inches). There are no known reports of adverse health consequences due to the missing pacemaker safety label or the increase of the minimum safety distance.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2517506-2018-00449
MDR Report Key7786204
Report SourceCOMPANY REPRESENTATIVE
Date Received2018-08-15
Date of Report2018-08-15
Date of Event2018-07-27
Date Mfgr Received2018-07-27
Date Added to Maude2018-08-15
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactKARL AEBIG
Manufacturer Street511 BENEDICT AVE
Manufacturer CityTARRYTOWN NY 10591
Manufacturer CountryUS
Manufacturer Postal10591
Manufacturer Phone9145243102
Manufacturer G1INPECO S.P.A
Manufacturer StreetVIA GIVOLETTO 15 ITALY REGISTRATION #: 30055092
Manufacturer City10040 VAL DELLA TORRE (TORINO),
Manufacturer CountryIT
Single Use3
Previous Use Code3
Removal Correction Number2517506-08/15/2018-007-C
Event Type3
Type of Report3

Device Details

Brand NameSTREAMLAB
Generic NameSTREAMLAB
Product CodeLXG
Date Received2018-08-15
Model NumberSTREAMLAB
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerSIEMENS HEALTHCARE DIAGNOSTICS INC.
Manufacturer Address500 GBC DRIVE PO BOX 6101 NEWARK DE 197146101 US 197146101


Patients

Patient NumberTreatmentOutcomeDate
10 2018-08-15

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.