KAIRISON PUNCH BRACKET PRESSURE REDUCER FK887R

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2018-08-16 for KAIRISON PUNCH BRACKET PRESSURE REDUCER FK887R manufactured by Aesculap Ag.

Event Text Entries

[117432710] Investigation: the investigation was carried out visually and a functional test was performed. The connectin pins are no longer according to the specifications, since they are deformed, most likely caused by a drop. Due to these damages, a proper mounting of the disposable tube set fk888su can no longer be guaranteed. The functional test was carried out successfully, but the set could only be assembled with high forces. Without a correct connection of the tube set, the function of the kairison bone punch will not be according to specification. Conclusion and root cause: based on the information available as well as a result of our investigation, the root cause of the failure is most probably related to an insufficient usage. No capa is necessary.
Patient Sequence No: 1, Text Type: N, H10


[117432711] Country of complaint: (b)(6). Io would not engage shaft.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9610612-2018-00347
MDR Report Key7790339
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2018-08-16
Date of Report2018-08-16
Date of Event2018-07-06
Date Facility Aware2018-07-16
Date Mfgr Received2018-08-16
Device Manufacturer Date2008-10-14
Date Added to Maude2018-08-16
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. NICOLE BROYLES
Manufacturer Street615 LAMBERT POINTE DRIVE
Manufacturer CityHAZELWOOD MO 63042
Manufacturer CountryUS
Manufacturer Postal63042
Manufacturer G1AESCULAP AG
Manufacturer StreetPO BOX 40
Manufacturer CityTUTTLINGEN, 78501
Manufacturer CountryGM
Manufacturer Postal Code78501
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameKAIRISON PUNCH BRACKET PRESSURE REDUCER
Generic NameRONGEUR, POWERED
Product CodeHAD
Date Received2018-08-16
Returned To Mfg2018-07-26
Model NumberFK887R
Catalog NumberFK887R
Lot Number51504186
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerAESCULAP AG
Manufacturer AddressPO BOX 40 TUTTLINGEN, 78501 GM 78501


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2018-08-16

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.