BENZ BENZODIAZEPINES PLUS 04490789190

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2018-08-17 for BENZ BENZODIAZEPINES PLUS 04490789190 manufactured by Roche Diagnostics.

Event Text Entries

[117446516] Na.
Patient Sequence No: 1, Text Type: N, H10


[117446517] The customer complained of false negative benz benzodiazepines plus results for 4 patient urine samples on a cobas 6000 c (501) module serial number (b)(4). Of the data provided, there was a discrepant benz result for 1 patient urine sample. The benzodiazepines result was negative on the c 501 module. The physician questioned the result and requested confirmation testing as the patient was taking clonazepam. The aminoclonazepam result was 156 ng/ml with the medmatch mass spectrophotometry using the same sample. According to product labeling, a 7-aminoclonazepam result of 144 ng/ml or greater should yield a positive result. There was no allegation of an adverse event. The product labeling states "many benzodiazepines appear in the urine largely as the glucuronidated conjugate. Glucuronidated metabolites may have more or less cross-reactivity than the parent compound. " the customer declined a service visit. The investigation was unable to find a definitive root cause.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1823260-2018-02746
MDR Report Key7792458
Report SourceHEALTH PROFESSIONAL,USER FACI
Date Received2018-08-17
Date of Report2018-08-17
Date of Event2018-07-26
Date Mfgr Received2018-07-26
Date Added to Maude2018-08-17
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1ROCHE DIAGNOSTICS GMBH
Manufacturer StreetSANDHOFERSTRASSE 116 NA
Manufacturer CityMANNHEIM (BADEN-WURTTEMBERG) 68305
Manufacturer CountryGM
Manufacturer Postal Code68305
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameBENZ BENZODIAZEPINES PLUS
Generic NameENZYME IMMUNOASSAY, BENZODIAZEPINE
Product CodeJXM
Date Received2018-08-17
Model NumberNA
Catalog Number04490789190
Lot Number279762
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457


Patients

Patient NumberTreatmentOutcomeDate
10 2018-08-17