HUDSON ISO-GARD HEPA,SMALL STRAIGHT 28062

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-08-17 for HUDSON ISO-GARD HEPA,SMALL STRAIGHT 28062 manufactured by Teleflex Medical.

Event Text Entries

[117833379] (b)(4). No oxygen saturation numerical value was reported. The device involved has not been received for evaluation at the time of this report. The investigation is still in progress.
Patient Sequence No: 1, Text Type: N, H10


[117833380] Customer complaint alleges that "the patient had a surgery under general anesthesia. At the end of the intervention, the ventilatory pressures increased suddenly, so the ventilation became impossible". Clinical consequence reported: the patient desaturated 30% during 1 {one} minute. There was a lack of "tital volumne," no capnography. A bronchospasm was considered. The patient was extubated after surgery and the patient is fine now. Interventions reported: deeping of anesthesia and ventolin did not resolve the issue. The intubation was changed but with no success. The filter was then replaced which resulted in the return of proper ventilation of the patient and quick re-saturation. Customer concluded "overall, the increase of the pressures was induced by the completely clogged antibacterial filter. The occlusion is due because of the moisture between the patient and the filter. "
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number8040412-2018-00213
MDR Report Key7794150
Date Received2018-08-17
Date of Report2018-08-02
Date of Event2018-07-31
Date Mfgr Received2018-09-21
Date Added to Maude2018-08-17
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactKATHARINE TARPLEY
Manufacturer Street3015 CARRINGTON MILL BLVD
Manufacturer CityMORRISVILLE NC 27560
Manufacturer CountryUS
Manufacturer Postal27560
Manufacturer Phone9194334854
Manufacturer G1TELEFLEX MEDICAL
Manufacturer StreetPO BOX 28, KAMUNTING INDUSTRIAL ESTATE
Manufacturer CityPERAK, WEST MALAYSIA 34600
Manufacturer CountryMY
Manufacturer Postal Code34600
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameHUDSON ISO-GARD HEPA,SMALL STRAIGHT
Product CodeCAH
Date Received2018-08-17
Returned To Mfg2018-09-06
Catalog Number28062
Lot NumberUNKNOWN
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerTELEFLEX MEDICAL
Manufacturer AddressPERAK, WEST MALAYSIA


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2018-08-17

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.