VAC-PAC 51630

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a user facility report with the FDA on 2018-08-17 for VAC-PAC 51630 manufactured by Natus Medical Incorporated.

Event Text Entries

[117857946] The customer was provided information for storage, repair, testing and replacement. Natus technical service advised the user that the vac-pac ifu notes to choose a vac-pac that supports the patient and to use additional supports as needed. Users are instructed to check the vac-pac device before and after each use and not to use the device if there is known damage or a leak. Users are also instructed to replace units older than two years.
Patient Sequence No: 1, Text Type: N, H10


[117857947] It was reported to natus that a vac-pac lost suction during a surgical procedure while connected to continuous suction. Upon loss of suction, the patient began sliding off of the operating room table. The anesthesiologist was able to stop the patient's slide. There was no harm caused to the patient. It was not reported whether there was any visible damage to the vac-pac. The vac-pac was reported to have been purchased over two years prior to the date of the event.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3018859-2018-00145
MDR Report Key7794970
Report SourceUSER FACILITY
Date Received2018-08-17
Date of Report2016-07-19
Date Mfgr Received2016-07-19
Date Added to Maude2018-08-17
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactJAMES FITZGERALD
Manufacturer Street5900 FIRST AVENUE SOUTH
Manufacturer CitySEATTLE WA 98108
Manufacturer CountryUS
Manufacturer Postal98108
Manufacturer Phone2062685173
Manufacturer G1NATUS MEDICAL INCORPORATED
Manufacturer Street5900 FIRST AVENUE SOUTH
Manufacturer CitySEATTLE WA 98108
Manufacturer CountryUS
Manufacturer Postal Code98108
Single Use3
Remedial ActionRL
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameVAC-PAC
Generic NameVAC-PAC
Product CodeCCX
Date Received2018-08-17
Model Number51630
Lot Number301724-509E
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerNATUS MEDICAL INCORPORATED
Manufacturer Address5900 FIRST AVENUE SOUTH SEATTLE WA 98108 US 98108


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2018-08-17