ACCU-CHEK ? CONNECT APP 07250452001

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a consumer report with the FDA on 2018-08-20 for ACCU-CHEK ? CONNECT APP 07250452001 manufactured by Roche Diabetes Care, Inc..

Event Text Entries

[117613909] Caller alleges inaccurate bolus advice. No adverse event reported. Requested back up file and meter requested to be returned for evaluation.
Patient Sequence No: 1, Text Type: D, B5


[130207231] Because the connect app is downloaded by the customer and installed on the customer's phone, it is not possible to get the actual device returned. The customer did return a back up file from their device and also returned the related blood glucose meter. Manufacturer was able to pair the meter with a retention version of the software and determine that the advice provided to the customer was correct. Manufacturer was also able to simulate results and transfer to the retention software and again all advice produced was correct.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3011393376-2018-03453
MDR Report Key7797885
Report SourceCONSUMER
Date Received2018-08-20
Date of Report2018-12-13
Date of Event2018-07-03
Date Mfgr Received2018-11-28
Date Added to Maude2018-08-20
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA GREG SMITH
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 462500457
Manufacturer CountryUS
Manufacturer Postal462500457
Manufacturer Phone3175212484
Manufacturer G1ROCHE DIABETES CARE, INC.
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 462501025
Manufacturer CountryUS
Manufacturer Postal Code462501025
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameACCU-CHEK ? CONNECT APP
Generic NameDIABETES MANAGEMENT SOFTWARE
Product CodeNDC
Date Received2018-08-20
Catalog Number07250452001
Lot Number2.1.4.1
OperatorLAY USER/PATIENT
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIABETES CARE, INC.
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457


Patients

Patient NumberTreatmentOutcomeDate
10 2018-08-20

© 2026 FDA.report