PATIENT ROOM WALL SAVER RECL 3500000730

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-08-20 for PATIENT ROOM WALL SAVER RECL 3500000730 manufactured by Stryker Medical-kalamazoo.

Event Text Entries

[117836332] It was reported that the wall saver mechanisms have been bending, so that the latch will not hold and the recliner "pops open. " no patient was affected and no adverse consequence or clinically relevant delay in treatment was reported.
Patient Sequence No: 1, Text Type: D, B5


[119944915] Upon evaluation of the unit, the stryker service technician found that the footrest of the unit was stuck extended, and could not be lowered. Therefore, the alleged issue of the footrest not latching and popping open was not confirmed. The issue of a stuck footrest, unable to be lowered, is not likely to cause or contribute to serious injury or death if it was to recur and is, therefore, not reportable.
Patient Sequence No: 1, Text Type: N, H10


[119944916] It was reported that the wall saver mechanisms have been bending, so that the latch will not hold and the recliner "pops open. " no patient was affected and no adverse consequence or clinically relevant delay in treatment was reported.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number0001831750-2018-00984
MDR Report Key7798172
Date Received2018-08-20
Date of Report2018-09-05
Date of Event2018-07-19
Date Mfgr Received2018-07-25
Device Manufacturer Date2013-11-11
Date Added to Maude2018-08-20
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. MARY KLAVER
Manufacturer Street3800 EAST CENTRE AVENUE
Manufacturer CityPORTAGE MI 49002
Manufacturer CountryUS
Manufacturer Postal49002
Manufacturer Phone2693292100
Manufacturer G1STRYKER MEDICAL-KALAMAZOO
Manufacturer Street3800 EAST CENTRE AVENUE
Manufacturer CityPORTAGE MI 49002
Manufacturer CountryUS
Manufacturer Postal Code49002
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NamePATIENT ROOM WALL SAVER RECL
Generic NameCHAIR, EXAMINATION AND TREATMENT
Product CodeFRK
Date Received2018-08-20
Catalog Number3500000730
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSTRYKER MEDICAL-KALAMAZOO
Manufacturer Address3800 EAST CENTRE AVENUE PORTAGE MI 49002 US 49002


Patients

Patient NumberTreatmentOutcomeDate
10 2018-08-20