CC CABLES 70CC2

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-08-20 for CC CABLES 70CC2 manufactured by Edwards Lifesciences.

Event Text Entries

[118097476] The cable is expected to be returned and be evaluated. A supplemental report will be submitted to communicate the device history record review results as well as the investigation results.
Patient Sequence No: 1, Text Type: N, H10


[118097477] It was reported that during use of the 70 cc2 cable, the blood temperature (bt) was unstable from 48? Celsius to 10? Celsius. When the suspect cable was replaced, the issue was improved. No alarm or error message was observed. No patient injury was reported and there was no treatment given to patient due to this event. Patient demographics are unavailable.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2015691-2018-03456
MDR Report Key7799405
Date Received2018-08-20
Date of Report2018-08-06
Date of Event2018-08-06
Date Mfgr Received2018-08-29
Device Manufacturer Date2009-11-18
Date Added to Maude2018-08-20
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. SAMANTHA EVELEIGH
Manufacturer StreetONE EDWARDS WAY
Manufacturer CityIRVINE CA 92614
Manufacturer CountryUS
Manufacturer Postal92614
Manufacturer Phone9492503939
Manufacturer G1EDWARDS LIFESCIENCES LLC
Manufacturer StreetONE EDWARDS WAY
Manufacturer CityIRVINE CA 92614
Manufacturer CountryUS
Manufacturer Postal Code92614
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameCC CABLES 70CC2
Generic NameCABLE
Product CodeISN
Date Received2018-08-20
Returned To Mfg2018-08-29
Model Number70CC2
Device Expiration Date2012-11-18
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerEDWARDS LIFESCIENCES
Manufacturer AddressONE EDWARDS WAY IRVINE CA 92614 US 92614


Patients

Patient NumberTreatmentOutcomeDate
10 2018-08-20

© 2026 FDA.report