SYMPHION FG-0612 74076

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2018-08-21 for SYMPHION FG-0612 74076 manufactured by Boston Scientific Corporation.

Event Text Entries

[117706219] It was reported to boston scientific corporation on (b)(6) 2018 that a symphion controller was used during a polypectomy procedure in the uterus performed on (b)(6) 2018. According to the complainant, during the procedure, upon connecting the resecting device into the symphion controller, the resecting device started resecting immediately without any prompt on the screen. The problem happened on three symphion resecting devices. The device was unplugged and it would not recognize the resecting device anymore that it was plugged in. The procedure was completed with another symphion controller. There were no patient serious injury or adverse effects to the patient as a result of the event.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3005099803-2018-60074
MDR Report Key7800703
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2018-08-21
Date of Report2018-09-19
Date of Event2018-08-08
Date Mfgr Received2018-08-28
Device Manufacturer Date2017-06-21
Date Added to Maude2018-08-21
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactCAROLE MORLEY
Manufacturer Street300 BOSTON SCIENTIFIC WAY
Manufacturer CityMARLBOROUGH MA 01752
Manufacturer CountryUS
Manufacturer Postal01752
Manufacturer Phone5086834015
Manufacturer G1NORTECH SYSTEMS
Manufacturer Street925 ENTERPRISE AVENUE ,
Manufacturer CityMILACA MN 56353
Manufacturer CountryUS
Manufacturer Postal Code56353
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSYMPHION
Generic NameINSUFFLATOR, HYSTEROSCOPIC, FLUID, CLOSED-LOOP RECIRCULATION WITH CUTTER-COAGULA
Product CodePGT
Date Received2018-08-21
Returned To Mfg2018-08-23
Model NumberFG-0612
Catalog Number74076
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBOSTON SCIENTIFIC CORPORATION
Manufacturer Address300 BOSTON SCIENTIFIC WAY MARLBOROUGH MA 01752 US 01752


Patients

Patient NumberTreatmentOutcomeDate
10 2018-08-21

© 2024 FDA.report