PKG, METZENBAUM SCISSORS, CURVED LONG 0250080265

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2018-08-21 for PKG, METZENBAUM SCISSORS, CURVED LONG 0250080265 manufactured by Stryker Endoscopy-san Jose.

Event Text Entries

[117778377] Mfg date: the device manufacture date is not known at this time. However, should it become available it will be provided in future reports. Additional information will be provided once the investigation has been completed.
Patient Sequence No: 1, Text Type: N, H10


[117778378] It was reported that a piece of the scissor fell off during a case. Additional information received states it was not known if the broken piece was retrieved from the patient.
Patient Sequence No: 1, Text Type: D, B5


[129639748] The product was not returned for investigation therefore the reported failure mode was not confirmed. The reported failure mode will be monitored for future reoccurrence. Alleged failure: broken. Probable root cause: poor autoclave reliability. Shear pin failure in handle. Use error. The device manufacture date is not known.
Patient Sequence No: 1, Text Type: N, H10


[129639769] It was reported that a piece of the scissor fell off during a case. Additional information received states it was not known if the broken piece was retrieved from the patient.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number0002936485-2018-00586
MDR Report Key7802398
Report SourceHEALTH PROFESSIONAL
Date Received2018-08-21
Date of Report2018-12-07
Date of Event2018-07-25
Date Mfgr Received2018-07-25
Date Added to Maude2018-08-21
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. CHESTER REBUGIO
Manufacturer Street5900 OPTICAL COURT
Manufacturer CitySAN JOSE CA 95138
Manufacturer CountryUS
Manufacturer Postal95138
Manufacturer Phone4087542000
Manufacturer G1STRYKER ENDOSCOPY-SAN JOSE
Manufacturer Street5900 OPTICAL COURT
Manufacturer CitySAN JOSE CA 95138
Manufacturer CountryUS
Manufacturer Postal Code95138
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NamePKG, METZENBAUM SCISSORS, CURVED LONG
Generic NameENDOSCOPIC GRASPING/CUTTING INSTRUMENT, NON-POWERED
Product CodeOCZ
Date Received2018-08-21
Catalog Number0250080265
Lot Number1644704
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerSTRYKER ENDOSCOPY-SAN JOSE
Manufacturer Address5900 OPTICAL COURT SAN JOSE CA 95138 US 95138


Patients

Patient NumberTreatmentOutcomeDate
10 2018-08-21