[118112993] The hill-rom technician found the external alarm was disconnected. The external alarm is the source of the bed exit alarm. It is necessary for the p500 to have an effective maintenance program. We recommend that you do annual preventive maintenance. Control unit checks:? The out of bed sensors operate correctly. A search of the hill-rom maintenance records showed hill-rom performed preventative maintenance on this bed in 2018. The technician reconnected the external alarm to resolve the issue. Based on this information, no further action is required.
Patient Sequence No: 1, Text Type: N, H10


[118112994] Hill-rom received a report from the account stating the alarm is not working. The mattress replacement system was located at the account. There was no patient/user injury reported. (b)(4).
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1824206-2018-00325
MDR Report Key7809019
Date Received2018-08-23
Date of Report2018-08-09
Date of Event2018-08-09
Date Mfgr Received2018-08-09
Device Manufacturer Date2011-04-13
Date Added to Maude2018-08-23
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactTONY WERNER
Manufacturer Street1069 STATE ROUTE 46 EAST
Manufacturer CityBATESVILLE IN 47006
Manufacturer CountryUS
Manufacturer Postal47006
Manufacturer Phone8129312359
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameP500 MRS CONTROL UNIT
Generic NameBED, FLOTATION THERAPY, POWERED
Product CodeIOQ
Date Received2018-08-23
Model NumberP005787RENT02
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerHILL-ROM BATESVILLE
Manufacturer Address1069 STATE ROUTE 46 EAST BATESVILLE IN 47006 US 47006


Patients

Patient NumberTreatmentOutcomeDate
10 2018-08-23

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.