COBAS AMPLICOR CT/NG TEST FOR C. TRACHOMATIS * 20757535122

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01 report with the FDA on 2006-11-15 for COBAS AMPLICOR CT/NG TEST FOR C. TRACHOMATIS * 20757535122 manufactured by Roche Molecular Systems, Inc..

Event Text Entries

[535362] Customer reported a period of specimens that were negative for chlamydia trachomatis with the roche molecular systems cobas amplicor ct/ng test, but positive with an alternate method.
Patient Sequence No: 1, Text Type: D, B5


[7914844] The customer provided roche molecular systems with samples from the specimen in question. Roche molecular systems, inc. Investigated the customer's allegations and confirmed that the specimens tested negative for chlamydia trachomatis. Further investigation determined that the specimens contained a mutated strain of c. Trachomatis. This mutation resulted in the deletion of approximately 350 base pairs from the region within the cryptic plasmid dna covered by the test's primers / and or probe. Rms is continuing our investigation to better understand the provalence, and geographic distribution of the mutated strain. The package insert will also be modified to reflect the rare, but possible occurrence of mutations that could impact the test's performance.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2243471-2006-00012
MDR Report Key781491
Report Source01
Date Received2006-11-15
Date of Report2006-10-13
Date Mfgr Received2006-10-20
Device Manufacturer Date1998-12-01
Date Added to Maude2006-11-17
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location0
Manufacturer ContactROBERT PIGOZZI
Manufacturer Street1080 US HIGHWAY 202
Manufacturer CityBRANCHBURG NJ 08876
Manufacturer CountryUS
Manufacturer Postal08876
Manufacturer Phone9257308272
Manufacturer G1*
Manufacturer Street*
Manufacturer City*
Manufacturer Country*
Single Use3
Remedial ActionNO
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCOBAS AMPLICOR CT/NG TEST FOR C. TRACHOMATIS
Generic NameIN VITRO DIAGNOSTIC NUCLEIC ACID TEST
Product CodeMKZ
Date Received2006-11-15
Model Number*
Catalog Number20757535122
Lot NumberNA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrY
Implant FlagN
Date Removed*
Device Sequence No1
Device Event Key769276
ManufacturerROCHE MOLECULAR SYSTEMS, INC.
Manufacturer Address* BRANCHBURG NJ 08876 US
Baseline Brand NameCOBAS AMPLICOR CT NG
Baseline Generic NameIN VITRO DIAGNOSTIC, NAT
Baseline Model No*
Baseline Catalog No20757535122
Baseline IDNA


Patients

Patient NumberTreatmentOutcomeDate
10 2006-11-15