ELECSYS PROLACTIN ASSAY 07027737190

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2018-08-27 for ELECSYS PROLACTIN ASSAY 07027737190 manufactured by Roche Diagnostics.

Event Text Entries

[118406749] This event occurred in (b)(6).
Patient Sequence No: 1, Text Type: N, H10


[118406770] The customer stated that they received an erroneous result for one patient sample tested with the elecsys prolactin assay on a cobas 8000 e 801 module. It was asked, but it is not known if the erroneous result was reported outside of the laboratory. The customer routinely tests samples for prolactin on their tosoh aia-2000 analyzer. The sample was initially tested on the aia-2000 analyzer, resulting with a prolactin value of 255. 8 ng/ml. The physician questioned the result and asked for the sample to be repeated on the e 801 analyzer. The repeat prolactin result from the e 801 analyzer was 343. 9 ng/ml. No adverse events were alleged to have occurred with the patient. The serial number of the e 801 analyzer was asked for, but not provided.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1823260-2018-02871
MDR Report Key7818081
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2018-08-27
Date of Report2018-10-29
Date of Event2018-07-23
Date Mfgr Received2018-08-08
Date Added to Maude2018-08-27
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1ROCHE DIAGNOSTICS GMBH
Manufacturer StreetSANDHOFERSTRASSE 116 NA
Manufacturer CityMANNHEIM (BADEN-WURTTEMBERG) 68305
Manufacturer CountryGM
Manufacturer Postal Code68305
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameELECSYS PROLACTIN ASSAY
Generic NameRADIOIMMUNOASSAY, PROLACTIN (LACTOGEN)
Product CodeCFT
Date Received2018-08-27
Model NumberNA
Catalog Number07027737190
Lot NumberASKU
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457