ELECSYS CORTISOL II 06687733190

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2018-08-27 for ELECSYS CORTISOL II 06687733190 manufactured by Roche Diagnostics.

Event Text Entries

[118562830] The event occurred in: (b)(6).
Patient Sequence No: 1, Text Type: N, H10


[118562831] The customer complained of an elecsys cortisol ii result for 1 patient sample on cobas 6000 e 601 module with serial number (b)(4). The initial cortisol result was 75. 1 nmol/l with a data flag and was released outside of the laboratory. The initial result did not match the patient's clinical picture so a second sample was collected. On (b)(6) 2018 the initial cortisol result for the second sample was 315. 5 nmol/l. The same sample was retested on (b)(6) 2018 with a cortisol result of 316. 6 nmol/l. There was no allegation of an adverse event. The investigation is currently ongoing.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1823260-2018-02884
MDR Report Key7819484
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2018-08-27
Date of Report2018-09-21
Date of Event2018-08-02
Date Mfgr Received2018-08-09
Date Added to Maude2018-08-27
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1ROCHE DIAGNOSTICS GMBH
Manufacturer StreetSANDHOFERSTRASSE 116 NA
Manufacturer CityMANNHEIM (BADEN-WURTTEMBERG) 68305
Manufacturer CountryGM
Manufacturer Postal Code68305
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameELECSYS CORTISOL II
Generic NameFLUOROMETRIC, CORTISOL
Product CodeJFT
Date Received2018-08-27
Model NumberNA
Catalog Number06687733190
Lot NumberASKU
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457


Patients

Patient NumberTreatmentOutcomeDate
10 2018-08-27