MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2018-08-27 for VITEK? 2 GRAM NEGATIVE ID TEST KIT 21341 manufactured by Biomerieux Inc..
[118892849]
On (b)(6) 2018, a customer in (b)(6) contacted biomerieux to report a misidentification of escherichia coli in association with the vitek? 2 gram-negative (gn) id test kit. The customer reported that vitek 2 identified an organism isolated from a urine specimen as shigella sonnei. The organism was a lactose fermenter. Shigella is a non-lactose fermenter and therefore, the isolate should not be a shigella organism. The customer sent the isolate to a reference lab. On 04-aug-2018, the reference laboratory reported that the organism was identified as escherichia coli. There is no indication or report from the laboratory that the discrepant result led to any adverse event related to the patient's state of health. A biom? Rieux internal investigation will be initiated.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 1950204-2018-00300 |
MDR Report Key | 7820362 |
Report Source | FOREIGN,HEALTH PROFESSIONAL,U |
Date Received | 2018-08-27 |
Date of Report | 2018-10-30 |
Date Mfgr Received | 2018-10-04 |
Device Manufacturer Date | 2018-03-22 |
Date Added to Maude | 2018-08-27 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Reporter Occupation | OTHER HEALTH CARE PROFESSIONAL |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | MRS. CANDACE MARTIN |
Manufacturer Street | 595 ANGLUM ROAD |
Manufacturer City | HAZELWOOD MO 63042 |
Manufacturer Country | US |
Manufacturer Postal | 63042 |
Manufacturer G1 | BIOMERIEUX INC. |
Manufacturer Street | 595 ANGLUM ROAD |
Manufacturer City | SAINT LOUIS MO 63042 |
Manufacturer Country | US |
Manufacturer Postal Code | 63042 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | VITEK? 2 GRAM NEGATIVE ID TEST KIT |
Generic Name | VITEK? 2 GN ID TEST KIT |
Product Code | LQM |
Date Received | 2018-08-27 |
Catalog Number | 21341 |
Lot Number | 2410481403 |
Device Expiration Date | 2019-03-22 |
Operator | HEALTH PROFESSIONAL |
Device Availability | N |
Device Age | DA |
Device Eval'ed by Mfgr | R |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | BIOMERIEUX INC. |
Manufacturer Address | 595 ANGLUM ROAD SAINT LOUIS MO 63042 US 63042 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2018-08-27 |