AERO T

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-08-27 for AERO T manufactured by Tisport, Llc.

Event Text Entries

[118552394] The injuries reported in this case were not considered to be serious, however there is a risk of serious injury in the case of a failure of anti-tips while engaged. The returned samples of the anti-tips were tested, along with samples from the then-current lot and of a different lot. The results of a combination of internal and external/third-party testing indicates that the material of the anti-tip tube used in the construction of this chair was softer and weaker than the design specified.
Patient Sequence No: 1, Text Type: N, H10


[118552395] The provider took the end user in their new wheelchair out to the parking lot of the hospital where the end user propelled themselves up a slight incline. Allegedly during this time, the chair rocked backwards and contacted the anti-tips three times, where on the third time the anti-tip tubes failed and the chair fell backwards. The provider claims that the end user hit his head and a ct was done, but that so far, he is okay.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3032618-2018-00004
MDR Report Key7820446
Date Received2018-08-27
Date of Report2018-08-27
Date of Event2018-07-30
Date Mfgr Received2018-07-30
Device Manufacturer Date2018-07-23
Date Added to Maude2018-08-27
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. BRIAN ENGLISH
Manufacturer Street2701 W COURT ST
Manufacturer CityPASCO WA 99301
Manufacturer CountryUS
Manufacturer Postal99301
Manufacturer Phone5095866117
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameAERO T
Generic NameWHEELCHAIR ANTI-TIP
Product CodeIMR
Date Received2018-08-27
Model NumberAERO T
OperatorLAY USER/PATIENT
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerTISPORT, LLC
Manufacturer Address2701 W COURT ST PASCO WA 99301 US 99301


Patients

Patient NumberTreatmentOutcomeDate
10 2018-08-27

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.