SYMBIONIC LEG 3 SMBE5260

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-08-27 for SYMBIONIC LEG 3 SMBE5260 manufactured by ?ssur Hf.

Event Text Entries

[118559072] The patient reported to have fallen while walking the stairs due to the knee, and indicated the electronics sometimes stop working without any reason. He is not injured. The product is currently at ossur quality center pending investigation. During the general quality check the leg did pass all the tests implying good functionality of the product. However, the error log showed that the electronic errors the user experienced during use indeed appeared, indicating malfunctioning of the leg might have contributed to the event. Issue at this moment seems to be a unique case. Issue will be further investigated to determine root cause and necessary actions. No initial corrective/preventive actions are considered warranted based on information currently available. There have been no similar incidents reported.
Patient Sequence No: 1, Text Type: N, H10


[118559073] The patient reported to have fallen while walking the stairs due to the knee, and indicated the electronics sometimes stop working without any reason. He is not injured.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3003764610-2018-00002
MDR Report Key7820748
Date Received2018-08-27
Date of Report2018-08-27
Date of Event2018-07-19
Date Mfgr Received2018-07-16
Date Added to Maude2018-08-27
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMRS. CRYSTAL SWAFFORD
Manufacturer Street27051 TOWNE CENTRE DRIVE
Manufacturer CityFOOTHILL RANCH CA 92610
Manufacturer CountryUS
Manufacturer Postal92610
Manufacturer Phone9493823817
Manufacturer G1OSSUR HF
Manufacturer StreetGRJOTHALS 1-5
Manufacturer CityREYKJAVIK, REYKJAVIK 110
Manufacturer CountryIC
Manufacturer Postal Code110
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameSYMBIONIC LEG 3
Generic NameSYMBIONIC LEG 3
Product CodeISW
Date Received2018-08-27
Returned To Mfg2018-08-17
Model NumberSMBE5260
Catalog NumberSMBE5260
Device AvailabilityR
Device Age3 YR
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
Manufacturer?SSUR HF
Manufacturer AddressGRJOTHALS 1-5 REYKJAVIK, REYKJAVIK 110 IC 110


Patients

Patient NumberTreatmentOutcomeDate
10 2018-08-27