EVIS EXERA II ULTRASONIC BRONCHOFIBERVIDEOSCOPE BF-UC180F

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-08-28 for EVIS EXERA II ULTRASONIC BRONCHOFIBERVIDEOSCOPE BF-UC180F manufactured by Olympus Medical Systems Corp..

Event Text Entries

[118546464] The device has not been returned to omsc for evaluation. Omsc reviewed the manufacturing history of this device and confirmed no irregularity. The exact cause of the reported event could not be conclusively determined at this time. If additional information is received, this report will be supplemented.
Patient Sequence No: 1, Text Type: N, H10


[118546465] Olympus medical systems corp. (omsc) was informed that as a result of microbiological testing by the user facility, mycobacterium lentiflavum were detected from the sample collected from the subject device. The number of microbes is unknown. The device had been reprocessed using an olympus automated endoscope reprocessor model mini etd2 (not available in the usa) with peracetic acid. The user facility reported that microbes were detected from three different devices including the subject device and there was a possibility that patients (up to 20 patients) who used these devices were infected. However, at this time there is no report of patient infection associated with this report. This is 2nd of 3 reports.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number8010047-2018-01657
MDR Report Key7821678
Date Received2018-08-28
Date of Report2018-09-21
Date Mfgr Received2018-08-28
Device Manufacturer Date2008-07-17
Date Added to Maude2018-08-28
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR KAZUTAKA MATSUMOTO
Manufacturer Street2951 ISHIKAWA-CHO
Manufacturer CityHACHIOJI-SHI, TOKYO-TO 192-8507
Manufacturer CountryJA
Manufacturer Postal192-8507
Manufacturer Phone426425177
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameEVIS EXERA II ULTRASONIC BRONCHOFIBERVIDEOSCOPE
Generic NameULTRASONIC BRONCHOFIBERVIDEOSCOPE
Product CodePSV
Date Received2018-08-28
Model NumberBF-UC180F
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerOLYMPUS MEDICAL SYSTEMS CORP.
Manufacturer Address2951 ISHIKAWA-CHO HACHIOJI-SHI, TOKYO-TO US


Patients

Patient NumberTreatmentOutcomeDate
10 2018-08-28