PRISM CHAGAS 07K35-68

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative report with the FDA on 2018-08-28 for PRISM CHAGAS 07K35-68 manufactured by Abbott Laboratories.

Event Text Entries

[118576184] Further investigation of the customer issue included a review of the complaint text, a search for similar complaints, review of instrument logs, a review of labeling, review of field data reactive rates, and a sensitivity and specificity review. No adverse trend was identified for the issue. Labeling was reviewed and found to be adequate. Both specificity and sensitivity reviews met all specifications. Based on all available information and abbott diagnostics' complaint investigation no product deficiency was identified.
Patient Sequence No: 1, Text Type: N, H10


[118576185] During a clinical evaluation study using prism chagas as the predicate product, a donor specimen tested on prism was found to be discordant relative to another supplemental method being used for final result interpretation. Two false non-reactive samples on the prism chagas assay for samples w14091712737300 and w14091801031634 were identified during the course of the study. The donor was deferred and the blood product was destroyed; there was no impact to patient management.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1415939-2018-00122
MDR Report Key7824143
Report SourceCOMPANY REPRESENTATIVE
Date Received2018-08-28
Date of Report2018-08-28
Date Mfgr Received2018-05-31
Device Manufacturer Date2017-09-01
Date Added to Maude2018-08-28
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNOEMI ROMERO-KONDOS, RN BSN
Manufacturer Street100 ABBOTT PARK ROAD DEPT. 09B9, LCCP1-3
Manufacturer CityABBOTT PARK IL 600643537
Manufacturer CountryUS
Manufacturer Postal600643537
Manufacturer Phone224667-512
Manufacturer G1ABBOTT LABORATORIES
Manufacturer Street100 ABBOTT PARK ROAD
Manufacturer CityABBOTT PARK IL 600643500
Manufacturer CountryUS
Manufacturer Postal Code600643500
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NamePRISM CHAGAS
Generic NameCHAGAS
Product CodeMIU
Date Received2018-08-28
Catalog Number07K35-68
Lot Number78049M500
Device Expiration Date2018-05-12
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerABBOTT LABORATORIES
Manufacturer Address100 ABBOTT PARK ROAD ABBOTT PARK IL 600643500 US 600643500


Patients

Patient NumberTreatmentOutcomeDate
10 2018-08-28