DUALPRO? IVUS+NIRS IMAGING CATHETER TVC-C195-42

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a distributor,foreign report with the FDA on 2018-08-29 for DUALPRO? IVUS+NIRS IMAGING CATHETER TVC-C195-42 manufactured by Infraredx, Inc..

Event Text Entries

[118557438] Percutaneous coronary intervention. Diagnostic imaging with a combination near-infrared spectroscopy (nirs) and intravascular ultrasound (ivus) catheter. A first imaging catheter was used to scan a coronary artery. A 90 degree-size wedge of the transverse ivus image was observed to be missing during the scan. A second catheter was tried and a missing wedge of the transverse ivus image was again observed during the scan. A different brand of ivus imaging catheter was then used to scan the vessel. A dissection of the vessel was observed during scanning and a stent was deployed to treat it. While the nirs-ivus catheter was not deemed to be the direct cause of patient harm, failure of the ivus imaging with the nirs-ivus catheter complicated the procedure and may have contributed to user error.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3004722468-2018-00002
MDR Report Key7826579
Report SourceDISTRIBUTOR,FOREIGN
Date Received2018-08-29
Date of Report2018-08-15
Date of Event2018-08-01
Date Facility Aware2018-08-08
Report Date2018-08-08
Date Reported to Mfgr2018-08-08
Date Mfgr Received2018-08-10
Device Manufacturer Date2018-04-23
Date Added to Maude2018-08-29
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. STEPHEN SUM
Manufacturer Street34 THIRD AVE.
Manufacturer CityBURLINGTON MA 018034414
Manufacturer CountryUS
Manufacturer Postal018034414
Manufacturer Phone7813459651
Manufacturer G1INFRAREDX, INC.
Manufacturer Street34 THIRD AVE.
Manufacturer CityBURLINGTON MA 018034414
Manufacturer CountryUS
Manufacturer Postal Code018034414
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameDUALPRO? IVUS+NIRS IMAGING CATHETER
Generic NameULTRASONIC PULSED ECHO IMAGING SYSTEM DIAGNOSTIC INTRAVASCULAR CATHETER
Product CodeOGZ
Date Received2018-08-29
Returned To Mfg2018-08-13
Model NumberTVC-C195-42
Lot Number804191
Device Expiration Date2018-10-19
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Age4 MO
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerINFRAREDX, INC.
Manufacturer Address34 THIRD AVE. BURLINGTON MA 018034414 US 018034414


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2018-08-29

© 2026 FDA.report