STA CLEANER SOLUTION 00973

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a distributor report with the FDA on 2018-08-29 for STA CLEANER SOLUTION 00973 manufactured by Diagnostica Stago S.a.s..

Event Text Entries

[118604475] (b)(4). Stago will provide a follow-up report(s) once relevant information has become available.
Patient Sequence No: 1, Text Type: N, H10


[118604476] On (b)(6) 2018: stago received a communication from (b)(6), the clinical laboratory educator/safety coordinator at (b)(6). She claims that they "received some sta cleaner solution for the evolution/compact that leaked because the box was crushed during transport" and when her inventory specialist was dumping it down the sink it splashed on him. It "burned his skin and discolored it". (b)(6) also stated that the shipping company took the reagent out of the original box, placed it into an unmarked box and wrapped it with packing tape. According to the person involved, "one of the packages was just a round ball of packing tape with the (b)(6) label taped on. It was not a box when opening. Everything was wet. The bottles were punctured and crushed. " at first, he did not know it was leaking since there was not any paperwork with the item, but realized what it was when he saw the bottles. Two of the 4 boxes shipped were damaged. On 16aug2018: (b)(6) was contacted by stago to obtain more information about the incident. According to her the person involved has a severe allergy to bleach. It should be noted that sta cleaner solution is a highly dilute disinfectant which does not contain bleach. Additionally, stago does not have any indication that gloves or other ppe were utilized. She stated, "he did not go to the er for further treatment, just self-treated a few blisters due to his allergy. Other than that all is ok. "
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9710666-2018-00001
MDR Report Key7827161
Report SourceDISTRIBUTOR
Date Received2018-08-29
Date of Report2018-12-20
Date of Event2018-08-15
Date Facility Aware2018-08-15
Report Date2018-08-24
Date Reported to Mfgr2018-08-24
Date Mfgr Received2018-08-17
Device Manufacturer Date2018-06-19
Date Added to Maude2018-08-29
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR WALID BEN AMMAR
Manufacturer Street2 RUE PIERRE FOSSATI
Manufacturer CityFRANCONVILLE VAL-DOISE, 95130
Manufacturer CountryFR
Manufacturer Postal95130
Manufacturer G1DIAGNOSTICA STAGO S.A.S.
Manufacturer Street3 ALLEE THERESA
Manufacturer CityASNIERE SUR SEINE, 92600
Manufacturer CountryFR
Manufacturer Postal Code92600
Single Use3
Remedial ActionRL
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSTA CLEANER SOLUTION
Generic NameSTA CLEANER SOLUTION
Product CodePPM
Date Received2018-08-29
Catalog Number00973
Lot Number202430
Device AvailabilityN
Device Age2 MO
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerDIAGNOSTICA STAGO S.A.S.
Manufacturer Address3 ALLEE THERESA ASNIERES SUR SEINE, 92600 FR 92600


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2018-08-29

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.