MAUDE MDR 7834697

MDR report key
7834697
Report number
1319808-2018-00024
Event key
0
Event type
3
Date of event
2018-08-02
Date received
2018-08-30
Adverse event
3
Product problem
3
Patients in event
0
Reporter occupation
0
Health professional
3
Initial report to FDA
3
Event location
3

Manufacturer Contact#

Contact
MR. JAMES A STEVENS
Address
100 INDIGO CREEK DRIVE ROCHESTER NY 14626 US
Phone
585-585-5854
Report source
M
Manufacturer link flag
Y

Devices#

Seq, Brand, Generic table
SeqBrandGenericManufacturerProduct codeModelCatalogLotPMA510(k)ImplantEvaluatedAvailability
1VITROS CHEMISTRY PRODUCTS VALP REAGENTIN VITRO DIAGNOSTICSORTHO-CLINICAL DIAGNOSTICSLEG68017102511-25-6393Y Y

Patients#

Sequence, Received, Treatment table
SequenceReceivedTreatmentOutcome
12018-08-300

Event Narratives#

N

Patient 1

THE INVESTIGATION DETERMINED THAT A LOWER THAN EXPECTED VITROS VALP RESULT WAS OBTAINED FROM A PATIENT SAMPLE PROCESSED AS PART OF A VALP LOT TO LOT CORRELATION USING A VITROS 5600 INTEGRATED SYSTEM. THE DEFINITIVE ASSIGNABLE CAUSE OF THIS EVENT COULD NOT BE DETERMINED. HOWEVER, A POSSIBLE CONTRIBUTOR TO THE EVENT WAS SUSPECTED TO BE CALIBRATION TO CALIBRATION VARIABILITY. THERE WAS NO INDICATION OF A PERFORMANCE ISSUE WITH EITHER VALP LOT (2511-25-6393 AND 2511-26-6711) BASED ON QC PERFORMANCE; HOWEVER, THE ACCURACY OF RESULTS OBTAINED USING VALP LOT 2511-25-6393 WAS LOWER THAN THE ACCURACY OF RESULTS OBTAINED USING VALP LOT 2511-26-6711 AS SHOWN WITH THE PATIENT CORRELATION RESULTS. CONTINUAL TRACKING AND TRENDING OF COMPLAINTS HAS NOT IDENTIFIED ANY SIGNALS THAT WOULD INDICATE A POTENTIAL SYSTEMIC ISSUE WITH EITHER OF THE VITROS VALP REAGENT LOTS. AN INSTRUMENT RELATED ISSUE IS NOT A LIKELY CAUSE OF THE EVENT AS THE TWO VITROS 5600 SYSTEMS GENERATED REPRODUCIBLE RESULTS.

D

Patient 1

THE CUSTOMER OBTAINED A LOWER THAN EXPECTED VITROS VALP RESULT (66.41 UG/ML VERSUS EXPECTED 83.17, 83.53, 86.31 UG/ML) FROM A PATIENT SAMPLE PROCESSED USING THE VITROS 5600 INTEGRATED SYSTEM. BIASED RESULTS OF THE MAGNITUDE AND DIRECTION OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION IF THEY WERE TO OCCUR UNDETECTED. THE PATIENT SAMPLE RESULT IN QUESTION WAS OBTAINED AS PART OF VALP LOT TO LOT CORRELATION STUDY AND THEREFORE, NO PATIENT RESULTS WERE REPORTED OUTSIDE THE LABORATORY. THERE WAS NO REPORTED ALLEGATION OF PATIENT HARM AS A RESULT OF THIS EVENT. THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC. COMPLAINT NUMBER (B)(4).