MAJ-173 TRANSFORMER WM-T1 100V K10000289

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-09-03 for MAJ-173 TRANSFORMER WM-T1 100V K10000289 manufactured by Keymed (medical And Industrial Equipment) Ltd..

Event Text Entries

[119506874] The fuse and terminal in the transformer was burned and damaged. No report of injury to patient or user as this was discovered during a pre-service inspection. Olympus (b)(4) have requested that the transformer is returned for further investigation. Reported in an abundance of caution and further information will be provided within a follow-up report.
Patient Sequence No: 1, Text Type: N, H10


[119506875] Google translation: as a result of "fuse breakage" pointed out to sorc nagano as a repair request, the terminal part of the fuse was inspected and confirmed to be burned and damaged. The condition that the connection terminal of the fuse holder was also burned and damaged. Checking with the local fs personnel, they advised that no health damage occurred following a discovering of the malfunction in pre-service inspection. Health damage has not occurred but because the connection terminal of the fuse holder is burned out, it may fall under the item of "safety management information pocket guide", so it was reported to the medical safety division and the etq raised.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9611174-2018-00014
MDR Report Key7840763
Date Received2018-09-03
Date of Report2018-10-08
Date Mfgr Received2018-10-03
Device Manufacturer Date2008-09-04
Date Added to Maude2018-09-03
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR PETER MORCOS
Manufacturer StreetKEYMED HOUSE, STOCK ROAD
Manufacturer CitySOUTHEND ON SEA, ESSEX SS25QH
Manufacturer CountryUK
Manufacturer PostalSS25QH
Manufacturer Phone0441702616
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameMAJ-173 TRANSFORMER WM-T1 100V
Generic NameWM-*P1 SERIES OF ENDOSCOPY WORKSTATIONS
Product CodeFEM
Date Received2018-09-03
Model NumberK10000289
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerKEYMED (MEDICAL AND INDUSTRIAL EQUIPMENT) LTD.
Manufacturer AddressKEYMED HOUSE, STOCK ROAD SOUTHEND-ON-SEA, ESSEX US


Patients

Patient NumberTreatmentOutcomeDate
10 2018-09-03

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.