TG GPO-PAP 04657594190

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2018-09-04 for TG GPO-PAP 04657594190 manufactured by Roche Diagnostics.

Event Text Entries

[119669184] This event occurred in (b)(6).
Patient Sequence No: 1, Text Type: N, H10


[119669185] The customer complained of discrepant results for 1 patient sample tested for trigl triglycerides (trigl) on a cobas c 111. The initial trigl result was 9. 50 mmol/l with a data flag. The sample was diluted 1:10 and repeated with a result of 3. 01 mmol/l. The sample was repeated neat and the result was 8. 97 mmol/l with no data flag. The result of 8. 97 mmol/l was reported outside of the laboratory. The sample was very lipemic so the high result was believed by the technician. On (b)(6) 2018 a new sample was obtained from the patient and the trigl result was 0. 76 mmol/l. Based on the lower result, the original sample was repeated and the result was 3. 23 mmol/l and the sample looked less lipemic. The patient was fasting when both samples were obtained. The customer is confident there was no sample mismatch. There was no allegation that an adverse event occurred. The c 111 serial number was (b)(4).
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1823260-2018-02947
MDR Report Key7841865
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2018-09-04
Date of Report2018-09-07
Date of Event2018-07-09
Date Mfgr Received2018-08-14
Date Added to Maude2018-09-04
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1ROCHE DIAGNOSTICS GMBH
Manufacturer StreetSANDHOFERSTRASSE 116 NA
Manufacturer CityMANNHEIM (BADEN-WURTTEMBERG) 68305
Manufacturer CountryGM
Manufacturer Postal Code68305
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameTG GPO-PAP
Generic NameTRIGLYCERIDES GPO
Product CodeCDT
Date Received2018-09-04
Model NumberNA
Catalog Number04657594190
Lot Number31272001
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457


Patients

Patient NumberTreatmentOutcomeDate
10 2018-09-04

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.