ERGO JUST CART- ACCESSORY TO EEG MEDICAL DEVICES IP-EC-DT-CART

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-09-10 for ERGO JUST CART- ACCESSORY TO EEG MEDICAL DEVICES IP-EC-DT-CART manufactured by Natus Medical Inc. Dba Excle-tech Ltd (xltek).

Event Text Entries

[119930593] In compliance with 21cfr part820. 198 - quality system regulations and natus quality management system, we initiated the investigation of the reported failure. The cause of the problem identified as deficiency in following the instruction. Customer installed a heavier monitor on the cart without adjusting the spring tension as explained in the user manual provided with the cart. Customer has been informed to adjust the tension if non-natus supplied monitor to use.
Patient Sequence No: 1, Text Type: N, H10


[119930594] Ergojust desktop cart - the adjustable work surface on the cart collapsed and hit someone on the knee causing brief pain on impact. The defective cart was taken out of use.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9612330-2018-00006
MDR Report Key7861895
Date Received2018-09-10
Date of Report2018-05-08
Date of Event2018-05-08
Date Mfgr Received2018-05-08
Date Added to Maude2018-09-10
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationBIOMEDICAL ENGINEER
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR SANJAY MEHTA
Manufacturer Street2568 BRISTOL CIRCLE
Manufacturer CityOAKVILLE, ONTARIO L6H5S1
Manufacturer CountryCA
Manufacturer PostalL6H5S1
Manufacturer G1NATUS MEDICAL INCORPORATED DBA EXCEL-TECH (XLTEK)
Manufacturer Street2568 BRISTOL CIRCLE
Manufacturer CityOAKVILLE, ONTARIO L6H5S1
Manufacturer CountryCA
Manufacturer Postal CodeL6H5S1
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameERGO JUST CART- ACCESSORY TO EEG MEDICAL DEVICES
Generic NameELECTROENCEPHALOGRAPH
Product CodeGWQ
Date Received2018-09-10
Returned To Mfg2018-06-12
Model NumberIP-EC-DT-CART
Lot NumberN/A
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerNATUS MEDICAL INC. DBA EXCLE-TECH LTD (XLTEK)
Manufacturer Address2568 BRISTOL CIRCLE OAKVILLE, OAKVILLE ONTARIO L6H 5S1 CA L6H 5S1


Patients

Patient NumberTreatmentOutcomeDate
10 2018-09-10

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.