MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2005-11-09 for * manufactured by .
[7913423]
The instrument has been investigated. The field service engineer (fse) evaluated the instrument and found a burnt resistor on the dispense (disp) printed circuit board in the wash/dispense module. The board was replaced. The instrument was tested without further problem and returned to service.
Patient Sequence No: 1, Text Type: N, H10
Report Number | 2250051-2005-00528 |
MDR Report Key | 787106 |
Report Source | 05,06 |
Date Received | 2005-11-09 |
Date Mfgr Received | 2005-10-10 |
Date Added to Maude | 2006-11-29 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 0 |
Product Problem Flag | 0 |
Reprocessed and Reused Flag | 0 |
Health Professional | 0 |
Initial Report to FDA | 0 |
Report to FDA | 3 |
Event Location | 0 |
Manufacturer Contact | LAURA VELLUCCI, PH.D |
Manufacturer Street | 1001 US HWY 202 |
Manufacturer City | RARITAN NJ 08869 |
Manufacturer Country | US |
Manufacturer Postal | 08869 |
Manufacturer Phone | 9082188532 |
Manufacturer G1 | * |
Manufacturer Street | * |
Manufacturer City | * |
Manufacturer Country | * |
Single Use | 3 |
Remedial Action | RP |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | * |
Product Code | JTL |
Date Received | 2005-11-09 |
Device Eval'ed by Mfgr | Y |
Implant Flag | N |
Device Sequence No | 1 |
Device Event Key | 774864 |
Baseline Brand Name | ORTHO SUMMIT PROCESSOR |
Baseline Generic Name | SAMPLE PROCESSOR |
Baseline Model No | 7003015 |
Baseline Catalog No | 936480 |
Baseline ID | NA |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2005-11-09 |