30-70ML CELL FREEZE CRYOGENIC STORAGE CONTAINER CF-250

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-09-13 for 30-70ML CELL FREEZE CRYOGENIC STORAGE CONTAINER CF-250 manufactured by Charter Medical, Ltd..

Event Text Entries

[120576291] The returned bag was visually inspected; a tear in the film, approximately 1/8-inch in length, was observed in the junction area of two bag welds. The film edges of the tear had a stretch appearance as if the film had been pulled apart. The failure mode was unique from previous leak reports involving this product style in several ways: previous complaints have been attributed to user handling technique issues after the bags were frozen. This one occurred during the filling process. Failure modes from previous reports presented themselves as small pin holes or film cracks. The exact root cause was undetermined. However, a review of the bag production process indicated a single manual process step where an issue of this nature could possibly occur. The bag is produced with 2 separate layers of film, which are manually placed on tooling and welded. If one of the film layers is not manually placed in a flat orientation, there is a possibility the weld tooling could pinch the film near the port weld junction and result in a tear or hole. This is the location where the bag tear was observed on the returned sample. Since there have been no known previous reports of this nature this issue appears to be an isolated occurrence.
Patient Sequence No: 1, Text Type: N, H10


[120576292] The end user experienced a leak from a possible film tear in a cf-250 device while the bag was being filled. The cell material was not salvaged from the bag and was not infused into the patient. There were no adverse events or additional medical intervention required as a result of this occurrence.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1066733-2018-00015
MDR Report Key7872603
Date Received2018-09-13
Date of Report2018-09-13
Date of Event2018-08-13
Date Mfgr Received2018-08-13
Device Manufacturer Date2017-11-08
Date Added to Maude2018-09-13
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. JESSICA HUGHES
Manufacturer Street3948-A WESTPOINT BLVD.
Manufacturer CityWINSTON SALEM 27103
Manufacturer CountryUS
Manufacturer Postal27103
Manufacturer Phone3367686447
Manufacturer G1CHARTER MEDICAL, LTD.
Manufacturer Street3948-A WESTPOINT BLVD.
Manufacturer CityWINSTON SALEM 27103
Manufacturer CountryUS
Manufacturer Postal Code27103
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand Name30-70ML CELL FREEZE CRYOGENIC STORAGE CONTAINER
Generic Name30-70ML CELL FREEZE CRYOGENIC STORAGE CONTAINER
Product CodeLPZ
Date Received2018-09-13
Returned To Mfg2018-08-21
Model NumberCF-250
Catalog NumberCF-250
Lot Number151223
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerCHARTER MEDICAL, LTD.
Manufacturer Address3948-A WESTPOINT BLVD. WINSTON SALEM NC 27103 US 27103


Patients

Patient NumberTreatmentOutcomeDate
10 2018-09-13

© 2024 FDA.report