OLYMPIC COOL-CAP 005094

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-09-13 for OLYMPIC COOL-CAP 005094 manufactured by Natus Medical Incorporated.

Event Text Entries

[120256986] Natus technical service has requested return of the system. Natus technical service has sent questions to the complainant regarding the details of the event and the patient involved in the event. A response to these requests has not been received.
Patient Sequence No: 1, Text Type: N, H10


[120256987] It was reported that a cool-cap system's cap temperature was high as compared to the set temperature. A patient was being treated with the system when the event occurred, and the patient was moved to another cool-cap system. There was no death, serious injury, delay in treatment, or environmental/safety concerns. A service technician who reported the event performed a heat load test as described in the cool-cap service manual and functional test procedure. The technician claimed that the test demonstrated that the internal thermistors were faulty. However, based on photos provided by the technician, natus factory service noted that the heat load test was performed incorrectly. Natus factory service stated that testing the functionality of the cooling modules would also require performing a different test.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3018859-2018-00150
MDR Report Key7873604
Date Received2018-09-13
Date of Report2018-08-14
Date Mfgr Received2018-10-23
Device Manufacturer Date2011-03-25
Date Added to Maude2018-09-13
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactJAMES FITZGERALD
Manufacturer Street5900 FIRST AVENUE SOUTH
Manufacturer CitySEATTLE WA 98108
Manufacturer CountryUS
Manufacturer Postal98108
Manufacturer Phone2062685173
Manufacturer G1NATUS MEDICAL INCORPORATED
Manufacturer Street5900 FIRST AVENUE SOUTH
Manufacturer CitySEATTLE WA 98108
Manufacturer CountryUS
Manufacturer Postal Code98108
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameOLYMPIC COOL-CAP
Generic NameCOOL-CAP
Product CodeMXM
Date Received2018-09-13
Returned To Mfg2018-10-23
Model Number005094
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerNATUS MEDICAL INCORPORATED
Manufacturer Address5900 FIRST AVENUE SOUTH SEATTLE WA 98108 US 98108


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2018-09-13

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.