CAPTURE-R READY INDICATOR RED CELLS 0006428

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2018-09-13 for CAPTURE-R READY INDICATOR RED CELLS 0006428 manufactured by Immucor, Inc..

Event Text Entries

[120717825] The full number for the product in question is (b)(4) (1/27/2003). Immucor technical support used a remote electronic connection method on 28aug2018 to assess the instrument test well images in question, and determined that the instrument test well images in question were visually negative. The immucor laboratory tested retention product on 29aug2018, which performed as expected. The internal immucor record tracking number for this record is (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[120717826] On (b)(6) 2018, a customer site reported an unexpected negative antibody identification when using capture-r ready indicator red cells on a galileo echo instrument, when tested on (b)(6) 2018.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1034569-2018-00208
MDR Report Key7874323
Report SourceHEALTH PROFESSIONAL,USER FACI
Date Received2018-09-13
Date of Report2018-09-13
Date of Event2018-08-27
Date Mfgr Received2018-08-28
Date Added to Maude2018-09-13
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR HOWARD YOREK
Manufacturer Street3130 GATEWAY DRIVE
Manufacturer CityNORCROSS GA 300915625
Manufacturer CountryUS
Manufacturer Postal300915625
Manufacturer Phone7704412051
Manufacturer G1IMMUCOR, INC.
Manufacturer Street3130 GATEWAY DRIVE
Manufacturer CityNORCROSS GA 300915625
Manufacturer CountryUS
Manufacturer Postal Code300915625
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCAPTURE-R READY INDICATOR RED CELLS
Generic NameANTIGLOBULIN COATED RED CELLS
Product CodeKSF
Date Received2018-09-13
Catalog Number0006428
Lot Number221154
Device Expiration Date2018-09-16
Device AvailabilityY
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerIMMUCOR, INC.
Manufacturer Address3130 GATEWAY DRIVE NORCROSS GA 300915625 US 300915625


Patients

Patient NumberTreatmentOutcomeDate
10 2018-09-13

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.