30-70ML CELL FREEZE CRYOGENIC STORAGE CONTAINER/3 LEADS CF-250-3L

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-09-14 for 30-70ML CELL FREEZE CRYOGENIC STORAGE CONTAINER/3 LEADS CF-250-3L manufactured by Charter Medical, Ltd..

Event Text Entries

[120576334] A single, used bag was received and visually inspected by engineering staff. Inspection revealed a film tear under the right spike port. The tear was jagged and non-linear, approximately 3/4-inch long, and extended from below the tear cap port to the perimeter weld location. A small portion of the tear extended approximately 1/8-inch into the perimeter weld. A capa had previously been issued to investigate film breaks of this nature. A review of the capa investigation indicated failure modes of this type have been reproduced by introducing an air bubble inside the bag prior to freezing. An air bubble could rapidly expand during the thawing process, resulting in a film tear. The evidence observed from the returned sample indicates the likely root cause of the cracked film was the existence of an air bubble in the bag prior to freezing. Precautions for air bubbles inside the bag prior to freezing are addressed in the product ifu's.
Patient Sequence No: 1, Text Type: N, H10


[120576335] The bag was used to store and transport a manufactured product. The end user experienced a leak from a possible film tear in a cf-250-3l device while the bag was being thawed after frozen for 3-months. The bag contained a finished manufactured product used for a patient study point. The contents of the bag were not infused into a patient. There were no adverse events or additional medical intervention required as a result of this occurrence.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1066733-2018-00014
MDR Report Key7876191
Date Received2018-09-14
Date of Report2018-09-14
Date of Event2018-08-14
Date Mfgr Received2018-08-14
Device Manufacturer Date2017-12-07
Date Added to Maude2018-09-14
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. JESSICA HUGHES
Manufacturer Street3948-A WESTPOINT BLVD.
Manufacturer CityWINSTON SALEM NC 27103
Manufacturer CountryUS
Manufacturer Postal27103
Manufacturer Phone3367686447
Manufacturer G1CHARTER MEDICAL, LTD.
Manufacturer Street3948-A WESTPOINT BLVD.
Manufacturer CityWINSTON SALEM NC 27103
Manufacturer CountryUS
Manufacturer Postal Code27103
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand Name30-70ML CELL FREEZE CRYOGENIC STORAGE CONTAINER/3 LEADS
Generic Name30-70ML CELL FREEZE CRYOGENIC STORAGE CONTAINER/3 LEADS
Product CodeLPZ
Date Received2018-09-14
Returned To Mfg2018-08-21
Model NumberCF-250-3L
Catalog NumberCF-250-3L
Lot Number149037
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerCHARTER MEDICAL, LTD.
Manufacturer Address3948-A WESTPOINT BLVD. WINSTON SALEM NC 27103 US 27103


Patients

Patient NumberTreatmentOutcomeDate
10 2018-09-14

© 2026 FDA.report