AXLE INTERSPINOUS FUSION SYSTEM X060-0318 NA

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-09-14 for AXLE INTERSPINOUS FUSION SYSTEM X060-0318 NA manufactured by X-spine Systems, Inc..

Event Text Entries

[120540286] The complainant reported that a cross bar plate grip-lock inserter broke. The complainant confirmed that there were no patient complications associated with the incident. A visual inspection of the device confirmed the complainant's description of the damage. A dhr was performed and the device met all required specifications prior to being released to distribution. The inserter ratchet is intended to hold the inserter in a fixed position with the implant loaded. If the handles are compressed further than the implant and ratchet allow, the distal end of he left handle, holding the implant may break. The inserter malfunction could be replicated if excess force was applied to the handles of the inserter. The inserter ratchet is intended to hold the inserter in a fixed position with the implant loaded. If the handles are compressed further than the implant ratchet allows, the handle of the inserter may break.
Patient Sequence No: 1, Text Type: N, H10


[120540287] The complainant reported that a cross bar plate grip-lock inserter broke. The complainant confirmed that there were no patient complications associated with the incident.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3005031160-2018-00027
MDR Report Key7877290
Date Received2018-09-14
Date of Report2018-09-14
Date Mfgr Received2018-08-21
Device Manufacturer Date2014-10-06
Date Added to Maude2018-09-14
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. CASEY MING
Manufacturer Street664 CRUISER LANE
Manufacturer CityBELGRADE MT 59714
Manufacturer CountryUS
Manufacturer Postal59714
Manufacturer Phone4063880480
Manufacturer G1X-SPINE SYSTEMS, INC.
Manufacturer Street664 CRUISER LANE
Manufacturer CityBELGRADE MT 59714
Manufacturer CountryUS
Manufacturer Postal Code59714
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameAXLE INTERSPINOUS FUSION SYSTEM
Generic NameINTERSPINOUS PROCESS FIXATION SYSTEM
Product CodePEK
Date Received2018-09-14
Returned To Mfg2018-08-24
Model NumberX060-0318
Catalog NumberNA
Lot NumberI8365-0000R
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerX-SPINE SYSTEMS, INC.
Manufacturer Address664 CRUISER LANE BELGRADE MT 59714 US 59714


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2018-09-14

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.