3.2 MM BI LEAD TO IS-1 BI IPG - PERMANENT EXTENSION, ADAPTOR KIT 5866-40M

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative report with the FDA on 2018-09-14 for 3.2 MM BI LEAD TO IS-1 BI IPG - PERMANENT EXTENSION, ADAPTOR KIT 5866-40M manufactured by Medtronic, Inc..

Event Text Entries

[120294087] Concomitant medical products: d224vrc, implanted: (b)(6) 2010; 6966110, implanted: (b)(6) 1994; 6961, implanted: (b)(6) 1994; 6707-15, implanted: (b)(6) 2010; 5866-40m, implanted: (b)(6) 2010. If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10


[120294088] It was reported that during laser lead extraction for a system upgrade, the patient required open heart surgery to repair a small tear in the right ventricle. A new system was implanted a few days later. No patient complications have been reported as a result of this event.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2182208-2018-01692
MDR Report Key7877348
Report SourceCOMPANY REPRESENTATIVE
Date Received2018-09-14
Date of Report2018-09-14
Date of Event2018-08-09
Date Mfgr Received2018-08-17
Device Manufacturer Date2009-04-09
Date Added to Maude2018-09-14
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactANNE SCHILLING
Manufacturer Street8200 CORAL SEA ST NE
Manufacturer CityMOUNDS VIEW MN 55112
Manufacturer CountryUS
Manufacturer Postal55112
Manufacturer Phone7635052036
Manufacturer G1MEDTRONIC, INC.
Manufacturer Street8200 CORAL SEA STREET NE
Manufacturer CityMOUNDS VIEW MN 55112
Manufacturer CountryUS
Manufacturer Postal Code55112
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand Name3.2 MM BI LEAD TO IS-1 BI IPG - PERMANENT EXTENSION, ADAPTOR KIT
Generic NameADAPTOR, LEAD, PACEMAKER
Product CodeDTD
Date Received2018-09-14
Model Number5866-40M
Catalog Number5866-40M
Device Expiration Date2011-03-14
OperatorLAY USER/PATIENT
Device AvailabilityN
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerMEDTRONIC, INC.
Manufacturer Address8200 CORAL SEA STREET NE MOUNDS VIEW MN 55112 US 55112


Patients

Patient NumberTreatmentOutcomeDate
101. Hospitalization; 2. Life Threatening; 3. Required No Informationntervention 2018-09-14

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.