ROCHE 9180 ELECTROLYTE ANALYZER ISE 9180

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2018-09-14 for ROCHE 9180 ELECTROLYTE ANALYZER ISE 9180 manufactured by Roche Diagnostics.

Event Text Entries

[120398827] This event occurred in (b)(6).
Patient Sequence No: 1, Text Type: N, H10


[120398828] The customer complained of discrepant ise indirect na for gen. 2 results for 1 patient sample on a roche 9180 electrolyte analyzer compared to an elyte instrument. The na result was 132 mmol/l on the 9180 electrolyte analyzer. The na result was 141 mmol/l on the elyte instrument. There was no allegation of an adverse event. No erroneous results were reported outside of the laboratory. The customer performed daily maintenance and calibration. The customer repeated the sample. The repeat na result was 137 mmol/l on the 9180 electrolyte analyzer and 142 mmol/l on the elyte instrument. The na electrode lot number and expiration date were not provided.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1823260-2018-03149
MDR Report Key7877768
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2018-09-14
Date of Report2018-10-19
Date of Event2018-08-28
Date Mfgr Received2018-08-28
Date Added to Maude2018-09-14
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1ROCHE INSTRUMENT CENTER AG TEGIMENTA
Manufacturer StreetFORRENSTRASSE NA
Manufacturer CityROTKREUZ 6343
Manufacturer CountrySZ
Manufacturer Postal Code6343
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameROCHE 9180 ELECTROLYTE ANALYZER
Generic NameELECTROLYTE ANALYZER
Product CodeJFP
Date Received2018-09-14
Model NumberISE 9180
Catalog NumberASKU
Lot NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457


Patients

Patient NumberTreatmentOutcomeDate
10 2018-09-14

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.