HEA 1.2 BEADCHIP KIT, SLIDE 800-20202-96

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-09-14 for HEA 1.2 BEADCHIP KIT, SLIDE 800-20202-96 manufactured by Bioarray Solutions Ltd..

Event Text Entries

[120716315] Sequencing results interpretation: the existence of cytosine and thymine at the polymorphic site that is indicative of the kpa/kpb antigens in the kel gene represents a heterozygous kpa+b+ individual1. An interfering polymorphism was responsible for the kpb- results on hea beadchip carriers heab5939_g6, heac1464_4 and heac1464_5. This polymorphism has been previously identified as c. 846g>c (prior nucleotide numbering is c. 966g>c; rs8175993), no clinical associations could be found in published literature. Interfering mutations are listed as a limitation in the precisetype hea molecular test package insert. This sample was previously investigated on may 5, 2016 with identical results via this id program.
Patient Sequence No: 1, Text Type: N, H10


[120716316] The customer reported a possible discrepancy. The donor is kpb- using the bioarray hea molecular beadchip kit; serology results were kpb+.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3005967741-2018-00013
MDR Report Key7877978
Date Received2018-09-14
Date of Report2018-09-14
Date of Event2018-08-14
Date Mfgr Received2018-08-14
Device Manufacturer Date2015-08-06
Date Added to Maude2018-09-14
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. LUZ VILLAMIZAR
Manufacturer Street35 TECHNOLOGY DRIVE SUITE 100
Manufacturer CityWARREN NJ 07059
Manufacturer CountryUS
Manufacturer Postal07059
Manufacturer Phone9084449591
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameHEA 1.2 BEADCHIP KIT, SLIDE
Generic NameHEA 1.2 BEADCHIP KIT, SLIDE
Product CodePEP
Date Received2018-09-14
Model Number800-20202-96
Catalog Number800-20202-96
Lot Number16-23
Device Expiration Date2015-12-31
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBIOARRAY SOLUTIONS LTD.
Manufacturer Address35 TECHNOLOGY DRIVE SUITE 100 WARREN NJ 07059 US 07059


Patients

Patient NumberTreatmentOutcomeDate
10 2018-09-14

© 2026 FDA.report