MCKESSON BRANDED SCBI 73-SCS025

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a distributor,user facility report with the FDA on 2018-09-14 for MCKESSON BRANDED SCBI 73-SCS025 manufactured by Sps Medical.

Event Text Entries

[120322537] It was reported an employee received a few abrasions on her neck after glass shattered from a self-contained steam biological indicator (bi). It was reported the employee pulled the (bi) that was still warm to the touch from the autoclave after waiting twenty minutes to crack it before transferring the bi to the incubator. When she cracked the bi it shattered and it was reported a few shards of glass caused abrasions on her neck. It was reported the employee is a new user to this device and was not wearing the correct ppe. She did not seek medical attention and is reported to be fine. This is an isolated incident, there has been no other reported events. This complaint will continue to be monitored in the sps complaint system.
Patient Sequence No: 1, Text Type: N, H10


[120322538] It was reported an employee received a few abrasions on her neck after glass shattered from a self-contained steam biological indicator (bi).
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1319130-2018-00002
MDR Report Key7878208
Report SourceDISTRIBUTOR,USER FACILITY
Date Received2018-09-14
Date of Report2018-09-14
Date of Event2018-08-15
Date Mfgr Received2018-08-15
Date Added to Maude2018-09-14
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationRISK MANAGER
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactLAUREN JOHNSON
Manufacturer Street14605 28TH AVE N
Manufacturer CityPLYMOUTH MN 55447
Manufacturer CountryUS
Manufacturer Postal55447
Manufacturer Phone7635533349
Manufacturer G1SPS MEDICAL
Manufacturer Street31 WATER STREET #1
Manufacturer CityCUBA NY 14727
Manufacturer CountryUS
Manufacturer Postal Code14727
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameMCKESSON BRANDED SCBI
Generic NameBIOLOGICAL INDICATOR
Product CodeFRC
Date Received2018-09-14
Returned To Mfg2018-08-28
Model Number73-SCS025
Lot Number6743
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerSPS MEDICAL
Manufacturer Address31 WATER STREET #1 CUBA NY 14727 US 14727


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2018-09-14

© 2026 FDA.report