MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional report with the FDA on 2018-09-14 for ARCHITECT B12 07K61-35 manufactured by A.i.d.d Longford.
[120561511]
An evaluation is in process. A follow-up report will be submitted when the evaluation is complete. This report is being filed on an international product, list number 7k61-25 / 35 that has a similar product distributed in the us, list number 7k61-27 / 37. This issue was also reported under mdr number 1628664-2018-01883 for the analyzer.
Patient Sequence No: 1, Text Type: N, H10
[120561512]
The customer observed falsely elevated results while using architect b12 reagents. The following data was provided. Sid (b)(6) initial 298 pmol/l, repeat 221 pmol/l, sid (b)(6) initial 415 pmol/l, repeat 219 pmol/l, sid (b)(6) initial 230 pmol/l, repeat 157 pmol/l, sid (b)(6) initial 376 pmol/l, repeat 214 pmol/l, sid (b)(6) initial 370 pmol/l, repeat 221 pmol/l, sid (b)(6) initial 239 pmol/l, repeat 133 pmol/l. No impact to patient management was reported.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 3005094123-2018-00050 |
MDR Report Key | 7879520 |
Report Source | FOREIGN,HEALTH PROFESSIONAL |
Date Received | 2018-09-14 |
Date of Report | 2018-10-25 |
Date of Event | 2018-08-23 |
Date Mfgr Received | 2018-10-03 |
Date Added to Maude | 2018-09-14 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Reporter Occupation | OTHER HEALTH CARE PROFESSIONAL |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | NOEMI ROMERO-KONDOS, RN BSN |
Manufacturer Street | 100 ABBOTT PARK ROAD DEPT. 09B9, LCCP1-3 |
Manufacturer City | ABBOTT PARK IL 600643537 |
Manufacturer Country | US |
Manufacturer Postal | 600643537 |
Manufacturer Phone | 224667-512 |
Manufacturer G1 | A.I.D.D LONGFORD |
Manufacturer Street | LISNAMUCK CO. LONGFORD |
Manufacturer City | LONGFORD NA |
Manufacturer Country | EI |
Manufacturer Postal Code | NA |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | ARCHITECT B12 |
Generic Name | VITAMIN B12 |
Product Code | CDD |
Date Received | 2018-09-14 |
Catalog Number | 07K61-35 |
Lot Number | 88078UI00 |
Operator | HEALTH PROFESSIONAL |
Device Availability | * |
Device Age | DA |
Device Eval'ed by Mfgr | N |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | A.I.D.D LONGFORD |
Manufacturer Address | LISNAMUCK CO. LONGFORD LONGFORD NA EI NA |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2018-09-14 |