TWO IS-1 UNI LEADS TO IS-1 BI IPG ADAPTOR KIT 5866-38M

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2018-09-17 for TWO IS-1 UNI LEADS TO IS-1 BI IPG ADAPTOR KIT 5866-38M manufactured by Medtronic, Inc..

Event Text Entries

[120516139] Concomitant medical products: 5568-45 lead, implanted on (b)(6) 2001; 452353s lead, implanted on (b)(6) 1994; 6932-65 lead, implanted on (b)(6) 1996. If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10


[120516140] It was reported that high thresholds, high and undefined impedance and noise were noted on the right atrial (ra) leads. It was noted that a previously implanted adapter was fixed to two ra leads and an implantable cardioverter defibrillator (icd). During the change out of the icd, the leads were capped and replaced with one ra lead and the adapter was removed. When doing the battery change of the icd the physician tested the two ra leads and thresholds were within normal limits. Therefore it was concluded that the adapter may be the issue. No patient complications have been reported as a result of this event.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2182208-2018-01702
MDR Report Key7883536
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2018-09-17
Date of Report2018-09-17
Date of Event2018-08-25
Date Mfgr Received2018-08-25
Device Manufacturer Date2012-03-14
Date Added to Maude2018-09-17
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactANNE SCHILLING
Manufacturer Street8200 CORAL SEA ST NE
Manufacturer CityMOUNDS VIEW MN 55112
Manufacturer CountryUS
Manufacturer Postal55112
Manufacturer Phone7635052036
Manufacturer G1MEDTRONIC, INC.
Manufacturer Street8200 CORAL SEA STREET NE
Manufacturer CityMOUNDS VIEW MN 55112
Manufacturer CountryUS
Manufacturer Postal Code55112
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTWO IS-1 UNI LEADS TO IS-1 BI IPG ADAPTOR KIT
Generic NameADAPTOR, LEAD, PACEMAKER
Product CodeDTD
Date Received2018-09-17
Model Number5866-38M
Catalog Number5866-38M
Device Expiration Date2014-02-14
OperatorLAY USER/PATIENT
Device AvailabilityN
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerMEDTRONIC, INC.
Manufacturer Address8200 CORAL SEA STREET NE MOUNDS VIEW MN 55112 US 55112


Patients

Patient NumberTreatmentOutcomeDate
101. Hospitalization; 2. Required No Informationntervention 2018-09-17

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.