INFINITY DOCKING STATION 7265338 7265338E5300

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01 report with the FDA on 2006-11-17 for INFINITY DOCKING STATION 7265338 7265338E5300 manufactured by Drager Medical Systems.

Event Text Entries

[555169] It was communicated to us that while the customer was using our delta pt monitor that was connected to our ids (infinity docking station) the anesthesiologist slipped and hit the lower corner of the monitor with his arm. The monitor lifted off with the upper part of the docking station and resulted in the separation of the top cover section. (where the pt monitor is mounted) from the bottom section of the ids. No pt injury was reported.
Patient Sequence No: 1, Text Type: D, B5


[8053364] The mfr draeger medical systems, inc. Received the ids (infinity docking station) correctly assembled. A representative from the mechanical engineering group evaluated the associated ids unit by performing a simulated functional test by installing a pt monitor on the associated ids unit and applying force in a way to possibly separate the top cover from the ids base component. The unit was also disassembled for further analysis. An inspection of the integrity of the pem fasteners was performed. This involved the removal of the ids top cover assembly to expose the fasteners. Each fastener was inspected for proper insertion into the molded plastic and appeared to be fine. Then a turning force was applied to each fastener to determine if the pem nut component was securely in place and would not move (spin or pull out). All pem nut fasteners stayed in place, none moved or pulled out of the molded plastic boss. After we had performed the above stated testing, we received info from our local field office that the pem fasteners had been glued into the molded plastic bosses after the reported event had occurred. Due to the applied glue on the pem fasteners, it was not possible to determine the root cause for the reported separation.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1220063-2006-00025
MDR Report Key788459
Report Source01
Date Received2006-11-17
Date of Report2006-11-15
Date of Event2006-06-13
Date Mfgr Received2006-06-13
Date Added to Maude2006-12-04
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location0
Manufacturer ContactSTEVE NAUGLE
Manufacturer Street6 TECH DRIVE
Manufacturer CityANDOVER MA 01810
Manufacturer CountryUS
Manufacturer Postal01810
Manufacturer Phone9783798000
Manufacturer G1*
Manufacturer Street*
Manufacturer City*
Manufacturer Country*
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameINFINITY DOCKING STATION
Generic NamePOWER SUPPLY AND NETWORK COMMUNICATION DEVICE
Product CodeMPE
Date Received2006-11-17
Returned To Mfg2006-07-17
Model Number7265338
Catalog Number7265338E5300
Lot Number546576065
ID Number*
OperatorNO INFORMATION
Device AvailabilityR
Device Eval'ed by MfgrY
Implant FlagN
Date Removed*
Device Sequence No1
Device Event Key776165
ManufacturerDRAGER MEDICAL SYSTEMS
Manufacturer Address6 TECH DR ANDOVER MA 01810 US
Baseline Brand NameINFINITY DOCKING STATION
Baseline Generic NamePOWER SUPPLY AND NETWORK COMMUNICATION DEVICE
Baseline Model No7265338
Baseline Catalog No7265338E5300
Baseline ID*


Patients

Patient NumberTreatmentOutcomeDate
10 2006-11-17

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.