Event Text Entries

[120572810] Old for tp earlier this year, a recall notice was sent out for the draeger jm-103 and jm-105 jaundice meters for the reason that, "users have misinterpreted display messages that have resulted in serious injuries. Specifically, the jm-103 meter displays three blinking dashes (- - -) and the jm-105 meter displays dash-zero-dash (-0-) when the bilirubin level in the patient is higher than the maximum level of detection (>340? Mol or 20mg/dl). Measurement of high bilirubin indicates a need for immediate medical evaluation. Some users have interpreted the two display messages as indicating a "low" or "zero" value instead of high bilirubin levels. When this happens, treatment may be delayed or not offered, which could lead to brain damage and possibly death in some newborns/infants. The out of range display is visible; however, the interpretation of the reading is not intuitive or clear. Uncertainty about the out of range indication on the jaundice meter could cause a delay of treatment in a patient with hyperbilirubinemia. " draeger provided labels to be applied to the meter to remind the user of the meaning of the displays. Despite our hospital having placed those labels on our machines, we experienced misinterpretation by multiple nurses and nas of the "0" when using the jm-105 and it was interpreted as a malfunction when the display on the jm-103 read "---". We feel the displays need to be redesigned using something other than 0 or dashes to indicate the bilirubin level is too high.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number7884759
MDR Report Key7884759
Date Received2018-09-18
Date of Report2018-08-31
Date of Event2018-08-26
Report Date2018-09-05
Date Reported to FDA2018-09-05
Date Reported to Mfgr2018-09-18
Date Added to Maude2018-09-18
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationRISK MANAGER
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameDRAGER JAUNDICE METER, MODEL JM-105
Generic NameBILIRUBIN (TOTAL AND UNBOUND) IN THE NEONATE TEST SYSTEM
Product CodeMQM
Date Received2018-09-18
Model NumberMU20105
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerDRAEGER MEDICAL SYSTEMS, INC.
Manufacturer Address3135 QUARRY ROAD TELFORD PA 18969 US 18969


Patients

Patient NumberTreatmentOutcomeDate
10 2018-09-18