MTS DISPENSE 1.0ML MTS9612 MTS DISPENSER

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2006-11-17 for MTS DISPENSE 1.0ML MTS9612 MTS DISPENSER manufactured by Micro Typing Systems, Inc..

Event Text Entries

[17017006] The customer reported that the 1. 0 ml mts dispenser was leaking and not dispensing the correct amount of volume. It is unknown if the instrument was being used for patient testing at the time of the incident.
Patient Sequence No: 1, Text Type: D, B5


[17233530] The customer reported that the mts dispenser was leaking and not dispensing properly. It is unknown if the issue was observed during testing. Customer labeling instructs customer to perform qc prior to all testing involving mts gel cards. Variations in red blood cell concentration can markedly affect the sensitivity of test results. If red blood cell suspensions are too concentrated, they can give weaker results due to the increase in the antigen/antibody ratio. In addition, red blood cells may fail to completely migrate to the bottom of the microtube and could cause a false positive interpretation. When red blood cells are too low in concentration, they become difficult to visualize, and in extreme cases, a weak positive can fail to be detected. Erroneous results were not reported as a result of this incident. The customer issue was resolved via replacement of product. Although these mitigations exist if the incident were to recur undetected a possible erroneous result may occur. The issue was not identified during a calibration of qc. Therefore, incident is reportable.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1056600-2006-00260
MDR Report Key789115
Report Source06
Date Received2006-11-17
Date of Report2006-11-17
Date of Event2006-10-19
Date Mfgr Received2006-10-19
Date Added to Maude2006-12-05
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location0
Manufacturer ContactDAVID HAGEN, SITE MANAGER
Manufacturer Street1295 SW 29TH AVENUE
Manufacturer CityPOMPANO BEACH FL 33069
Manufacturer CountryUS
Manufacturer Postal33069
Manufacturer Phone9546239528
Manufacturer G1*
Manufacturer Street*
Manufacturer City*
Manufacturer Country*
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameMTS DISPENSE 1.0ML
Generic NameMTS DISPENSER 1.0ML
Product CodeGJG
Date Received2006-11-17
Model NumberMTS9612
Catalog NumberMTS DISPENSER
Lot Number*
ID Number*
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrR
Implant FlagN
Date Removed*
Device Sequence No1
Device Event Key776796
ManufacturerMICRO TYPING SYSTEMS, INC.
Manufacturer Address1295 S.W. 29TH AVE. POMPANO BEACH FL 33069 US
Baseline Brand NameMTS DISPENSE 1.0ML
Baseline Generic NameMTS DISPENSER 1.0ML
Baseline Model NoMTS9612
Baseline Catalog NoMTS DISPENSER
Baseline ID*


Patients

Patient NumberTreatmentOutcomeDate
10 2006-11-17

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.