MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2006-12-01 for CELL-DYN 22 CALIBRATOR 99120-01 manufactured by Abbott Diagnostics Division/celldyn.
[572804]
The customer requested replacement of cell-dyn 22 calibrator due to issues with several parameters. There is no report of suspect patient results.
Patient Sequence No: 1, Text Type: D, B5
[7806360]
On-market instability for platelets with cell-dyn 22 calibrator lots 3098 and 3099 is being experienced. The calibrator could read higher by at least 10% on the cell-dyn ruby and as much as 22% on the cell-dyn 3200. If this calibrator is used and the calibration factor on the instrument is adjusted, a negative platelet sample bias could result. As a precaution, lot 3100 was also removed from use. Investigation into the cause of this instability is in progress. This is an initial report. An investigation is in process. A final report will be submitted when the investigation is complete.
Patient Sequence No: 1, Text Type: N, H10
Report Number | 2919069-2006-00100 |
MDR Report Key | 789680 |
Report Source | 06 |
Date Received | 2006-12-01 |
Date of Report | 2006-11-07 |
Date Mfgr Received | 2006-11-07 |
Device Manufacturer Date | 2006-08-01 |
Date Added to Maude | 2006-12-06 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 0 |
Manufacturer Contact | DR. DAVID BARCH, D-09Y6, AP6C-2 |
Manufacturer Street | 100 ABBOTT PARK RD |
Manufacturer City | ABBOTT PARK IL 600646092 |
Manufacturer Country | US |
Manufacturer Postal | 600646092 |
Manufacturer Phone | 8479379328 |
Manufacturer G1 | * |
Manufacturer Street | * |
Manufacturer City | * |
Manufacturer Country | * |
Single Use | 3 |
Remedial Action | RC |
Previous Use Code | 3 |
Removal Correction Number | 2919069-9/26/06-003-C |
Event Type | 3 |
Type of Report | 3 |
Brand Name | CELL-DYN 22 CALIBRATOR |
Generic Name | CD22 CALIBRATOR |
Product Code | KRZ |
Date Received | 2006-12-01 |
Model Number | NA |
Catalog Number | 99120-01 |
Lot Number | 3100 |
ID Number | NA |
Device Expiration Date | 2006-10-06 |
Operator | UNKNOWN |
Device Availability | Y |
Device Eval'ed by Mfgr | Y |
Implant Flag | N |
Date Removed | * |
Device Sequence No | 1 |
Device Event Key | 777354 |
Manufacturer | ABBOTT DIAGNOSTICS DIVISION/CELLDYN |
Manufacturer Address | 5440 PATRICK HENRY DR. SANTA CLARA CA 95054 US |
Baseline Brand Name | CELL-DYN 22 CALIBRATOR |
Baseline Generic Name | CD22 CALIBRATOR |
Baseline Model No | NA |
Baseline Catalog No | 99120-01 |
Baseline ID | NA |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2006-12-01 |