FINESSE SHARKSKIN ILM FORCEPS 711.88

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-09-24 for FINESSE SHARKSKIN ILM FORCEPS 711.88 manufactured by Alcon Grieshaber Ag.

Event Text Entries

[121713097] No sample has been received by manufacturing for evaluation. The device history record for the affected lot was reviewed. No abnormalities that could have contributed to this event were found in the production documentation and the sample was released according to acceptance criteria. A 100% final inspection is performed for this product. Investigation including root cause analysis is in progress. A supplemental mdr will be filed as necessary in accordance with 21 cfr 803. 56 when additional reportable information becomes available. The manufacturer internal reference number is: (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[121713098] A physician reported that an ophthalmic forceps tip was unable to be opened after it was closed during a cataract and vitrectomy surgery. An alternate forceps device was obtained in order to complete the procedure. There was no impact to the patient. Additional information is not available for this report.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3003398873-2018-00028
MDR Report Key7901786
Date Received2018-09-24
Date of Report2018-11-20
Date of Event2018-09-11
Date Mfgr Received2018-11-06
Device Manufacturer Date2018-05-15
Date Added to Maude2018-09-24
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. BRYAN BLAKE
Manufacturer Street6201 SOUTH FREEWAY MAIL STOP AB2-6
Manufacturer CityFORT WORTH TX 76134
Manufacturer CountryUS
Manufacturer Postal76134
Manufacturer Phone8176152230
Manufacturer G1ALCON GRIESHABER AG
Manufacturer StreetWINKELRIEDSTRASSE 52
Manufacturer CitySCHAFFHAUSEN 8203
Manufacturer CountrySZ
Manufacturer Postal Code8203
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report0

Device Details

Brand NameFINESSE SHARKSKIN ILM FORCEPS
Generic NameFORCEPS, OPHTHALMIC
Product CodeHNF
Date Received2018-09-24
Model NumberNA
Catalog Number711.88
Lot NumberF163445
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerALCON GRIESHABER AG
Manufacturer AddressWINKELRIEDSTRASSE 52 SCHAFFHAUSEN 8203 SZ 8203


Patients

Patient NumberTreatmentOutcomeDate
10 2018-09-24