JMS WINGEATER SAFETY A-V FISTULA NEEDLE 15G X 1" BE W/CLAMP 820-5002-33

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-09-19 for JMS WINGEATER SAFETY A-V FISTULA NEEDLE 15G X 1" BE W/CLAMP 820-5002-33 manufactured by Jms Singapore.

Event Text Entries

[121586172] At 1707, 2 hours into a scheduled 3 hour hemodialysis treatment, pt was found unresponsive. Vs at 1655: b/p 119/66; pulse 85. When attempting to administer normal saline, it was noted the venous needle was dislodged, with approx 1. 5 l blood loss. Pressure was held to cannulation site, and normal saline administered via arterial line; 911 was called, cpr was initiated and aed applied which advised no shock. A total of 3 liters of normal saline was administered during the resuscitation. Pt was discharged unresponsive from the clinic and transported to the hosp via ems, where the pt died. Upon inspection of the venous avf needle, it was noted that tape and gauge dressing were still intact on the fistula needle. The machine was sequestered and a functional test was performed. All required parameters passed.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number7901867
MDR Report Key7901867
Date Received2018-09-19
Date of Report2018-09-19
Date of Event2018-09-06
Date Facility Aware2018-09-06
Report Date2018-09-19
Date Reported to FDA2018-09-19
Date Reported to Mfgr2018-09-19
Date Added to Maude2018-09-24
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag0
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Sequence Number: 1

Brand NameJMS WINGEATER SAFETY A-V FISTULA NEEDLE
Generic NameFISTULA NEEDLE
Product CodeFIE
Date Received2018-09-19
Model Number15G X 1" BE W/CLAMP
Catalog Number820-5002-33
Lot Number180427391
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrI
Device Sequence No1
Device Event Key0
ManufacturerJMS SINGAPORE
Manufacturer AddressSINGAPORE 569620 569620

Device Sequence Number: 2

Brand NameJMS WINGEATER SAFETY A-V FISTULA NEEDLE
Generic NameFISTULA NEEDLE
Product CodeFIE
Date Received2018-09-19
Model Number15G X 1" BE W/CLAMP
Catalog Number820-5002-33
Lot Number180427391
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrI
Device Sequence No2
Device Event Key0
ManufacturerJMS SINGAPORE
Manufacturer AddressSINGAPORE 569620 SN 569620


Patients

Patient NumberTreatmentOutcomeDate
101. Death; 2. Hospitalization 2018-09-19